Hds 1033-Lect 11 - Aec

download Hds 1033-Lect 11 - Aec

of 46

Transcript of Hds 1033-Lect 11 - Aec

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    1/46

    Automatic Exposure Control

    FAIRUZ BINTI MOHD NASIRRADIOGRAPHY , FPSK

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    2/46

    Lecture Overview Introduction on AEC

    Purpose of AEC

    Types of detectors Technical consideration with AEC

    Limitation of AEC

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    3/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    4/46

    Introduction The Automatic Exposure Control (AEC) requires a special

    understanding on the part of the radiologic technologist.

    AEC is one method for setting exposure factors to ensure thata quality radiograph image is produce.

    The AEC is a device that measures the quantity of radiation

    that reaches the image receptor.

    It automatically terminates the exposure when the imagereceptor has received the required radiation intensity.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    5/46

    Automatic exposure control devices can assist theradiographer in producing consistent radiographic imagesfrom patient to patient, regardless of size or presence of

    pathology.

    The advantages of this consistency are numerous andinclude:

    decreased repeat rate;

    decreased patient exposure; and

    increased department efficiency.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    6/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    7/46

    Purpose of AECAEC is a system used to consistently control

    radiographic density by terminating the length ofexposure based on the amount of radiation reachingthe image receptor.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    8/46

    When an AEC device is used to terminate anexposure, the technologist sets the kVp and mA,

    but the time of the exposure is automaticallydetermined by the machine.

    The AEC device differs from a manual timer

    because the AEC does not stop the exposureuntil the film has reached an appropriate density.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    9/46

    X Ray tube

    Collimator

    Beam

    Soft

    tissueBoneAir Patient

    Table

    Grid

    Cassette

    AEC detectors

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    10/46

    AEC should produce consistently optimal densityradiographs because based on sensitometric,specific amount of radiation to the film producespecific density.

    If the x-ray exposure is terminated when the

    exposure corresponding to optimal density isreached, the resultant radiograph shoulddemonstrate optimal density.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    11/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    12/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    13/46

    AEC SystemsAEC devices work by the same principle of operation:radiation is transmitted through the patient

    converted into an electrical signal

    terminating the radiographic exposure.

    Two types of AEC have been used:Phototimers

    Ionization chamber (most common)

    Regardless of the specific type of AEC system used,almost all systems use a set of three detectors, arrangein same specific manner.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    14/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    15/46

    Phototimer type AEC Units

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    16/46

    Phototimer Type AEC Units Consist of

    a fluorescent screen,

    a photomultiplier tube (PMT) or photo tube and a complex circuit connected to a timer designed to terminate the

    exposure.

    The fluorescent screen is adjacent to the photomultiplier

    tube and both lie behind the Bucky grid and cassette.

    considered as exit-type devices because the detectors

    are positioned behind the image receptor

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    17/46

    Phototimer Type AEC Units Light paddles coated with fluorescent material; converting

    radiation to the light.

    The PMT receives the light energy and converts it toelectrical energy

    The timer is tripped and the radiographic exposure isterminated when a sufficient large charge has been

    exposed

    The amount of light produced by the screen depends onthe amount and energy of the X-rays reaching thefluorescent screen.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    18/46

    Phototimer AEC

    X-RAY TABLE

    X-RAY TUBE

    CASSETTE

    SCREEN

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    19/46

    PM TUBELIGHT

    ELECTRONS

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    20/46

    Ionization chamber system

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    21/46

    Ionization chamber system Ionization chamber is a hollow cell that contains air

    and is connect to the timer circuit via electrical wire.

    Ionization chamber AEC devices are consideredentrance-type devices because the detector arepositioned in front of image receptor.

    Compare to phototimers, ion chambers are lessaccurate, but there are prone to failure.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    22/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    23/46

    IONIZATION CHAMBER AEC

    X-RAY TABLE

    X-RAY TUBE

    BUCKY

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    24/46

    X-rayspenetrate the

    thinaluminumshell and

    strike the airmolecules.

    Electronreleased

    andionizationoccurs.

    The electronsare attracted

    to thepositively

    chargedelectrode and

    are carriedthrough anelectrical

    circuit

    Thechargetravel

    alongthe wireto thetimer

    circuit.

    The timer istripped and the

    radiographicexposure isterminated

    when asufficient large

    charge hasbeen exposed

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    25/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    26/46

    Even though the ionization chamberand the phototimer operate

    differently, they both have thesame function:

    convert radiation into an electrical

    signal which will be used to

    automatically stop the exposure

    when the film has reached the

    proper density

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    27/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    28/46

    Technical Considerations with AEC

    To use AEC to its advantage,

    radiographers must be awareof some important technical

    consideration peculiar to AEC

    system.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    29/46

    FACTORSINFLUENCING

    AEC

    Patientpositioning

    Detectorselection

    Densitysetting

    CollimationBackuptimer

    Effects ofmA, kVand SID

    Buckyselection

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    30/46

    Patient Positioning AEC units use detectors to receive radiation passing

    through the patient, it is important to properlyposition the anatomical part of interest over the

    detector. If the detector receives radiation coming from

    another anatomical area, the exposure can be eithertoo long or too short, overexposing or underexposing

    the film. If the radiographer positions the anatomical part

    correctly, the AEC unit will produce a radiographexhibiting proper density.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    31/46

    Patient Positioning If the radiographer positions the anatomical part correctly, the

    AEC unit will produce a radiograph exhibiting proper density.

    Without a technologist who is very knowledgeableabout anatomy and positioning, automatic timers are

    worthless.

    In fact, they may actually decrease departmentefficiency because of the increased amount of repeatradiographs that will result if used improperly.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    32/46

    Detector Selection Proper detector selection must be made for an AEC unit to

    produce the desired radiographic density.

    There are usually 3 detectors that are connected to a

    complex electrical circuit.

    Any single detector or a combination of two or three

    detectors, can be selected for a radiographic exposure.

    Failure to select the proper detectors will increase patient

    dosage because the exposure must be repeated to correctthe underexposure or overexposure of he first radiograph.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    33/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    34/46

    Density Setting The AEC devices are equipped with density

    controls allow the radiographer to fine tune theradiographic density that is produce by the unit.

    These are generally are in the form of buttons on thecontrol panel that are numbered -2, -1, +1, +2.

    These button changes exposure by somepredetermined amount or increment expressed aspercentage.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    35/46

    +4

    +3

    +2

    +1

    0

    -1

    -2 -3

    -4

    +100

    +75

    +50

    +25

    0

    -25

    -50 -75

    -100

    D. SETTING % DENSITY CHANGE

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    36/46

    Backup Timer The backup timer is safety device that prevents

    excessive exposure to the patient in cases of

    timer switch failure or radiographer error.

    The backup timer must limit the exposure to a

    maximum of 600mAs

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    37/46

    Collimation Collimation affects the production of scatter radiation.

    As collimation increases, the field size decreases and the quantity ofradiation decreases.

    The detector is unable to distinguish transmitted radiation fromscattered radiation.

    Because the detector is measuring both type of radiation, the timer isturn off too soon when scattered is excessive, which result inunderexposure of area of interest.

    The radiographer should open the collimator to the extent that the partof being radiographed is image properly, but not so much as cause tostop the AEC device to stop the exposure before the area being image isproperly exposed.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    38/46

    Bucky selection The Bucky must be correctly selected.

    Failure to do so results in the patient and the imagereceptor being exposed to excessive radiation.

    The backup time is reached, the exposureterminated.

    The radiographer should be certain to deactive the

    AEC system and use manual technique whenperforming any radiographic study where the imagereceptor is located outside the Bucky.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    39/46

    Effects of mA, kV and SID When using the AEC units, the radiographer does not

    set the exposure time.

    But the radiographer can influence the exposure time

    according to the mA and kV setting he chooses. If the mA is changed, X-ray quantity also changes, high

    mA, produces more X-rays so the exposure time will beshorter and vice versa.

    kV setting also influences exposure time when using

    AEC. High kV setting, higher energy X-rays are produced,

    more of the X-rays will also penetrate the part ,resulting shorter required exposure time. Vice versa.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    40/46

    EFFECTS OF mA, kVp, SID ON AEC

    mA AEC TIME

    kVp AEC TIME

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    41/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    42/46

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    43/46

    Interchangeability of film-screen

    system Different film-screen systems cannot be interchanged

    easily once an AEC is calibrated to produce specificdensities.

    When calibration is performed, it is done for aparticular film-screen speed.

    The AEC device cannot sense when the radiographer

    use different film or screen, resulting in too much ortoo little density.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    44/46

    Minimum Response Time The minimum response time represents the shortest

    exposure time possible.

    Minimum response time usually longer with AECsystem.

    This can be problem with some segment of patientpopulation, such as pediatric patient and

    uncooperative patients.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    45/46

    Lack of calibration Failure to maintain regular calibration of the unit and

    AEC device result in radiographs that lack consistence,reproducible, and appropriate density.

  • 8/3/2019 Hds 1033-Lect 11 - Aec

    46/46