PONTIFÍCIA UNIVERSIDADE CATÓLICA DO PARANÁ ÁREA DE...

63
PONTIFÍCIA UNIVERSIDADE CATÓLICA DO PARANÁ CENTRO DE CIÊNCIAS BIOLÓGICAS E DA SAÚDE PROGRAMA DE PÓS-GRADUAÇÃO EM ODONTOLOGIA ÁREA DE CONCENTRAÇÃO ORTODONTIA LUÍS FILIPE SIU LON RESISTÊNCIA ADESIVA AO CISALHAMENTO DE BRACKETS CERÂMICOS MONOCRISTALINOS: ESTUDO IN VITRO CURITIBA 2008

Transcript of PONTIFÍCIA UNIVERSIDADE CATÓLICA DO PARANÁ ÁREA DE...

  • PONTIFÍCIA UNIVERSIDADE CATÓLICA DO PARANÁ

    CENTRO DE CIÊNCIAS BIOLÓGICAS E DA SAÚDE

    PROGRAMA DE PÓS-GRADUAÇÃO EM ODONTOLOGIA

    ÁREA DE CONCENTRAÇÃO ORTODONTIA

    LUÍS FILIPE SIU LON

    RESISTÊNCIA ADESIVA AO CISALHAMENTO DE BRACKETS CERÂMICOS MONOCRISTALINOS: ESTUDO IN VITRO

    CURITIBA

    2008

  • Livros Grátis

    http://www.livrosgratis.com.br

    Milhares de livros grátis para download.

  • LUÍS FILIPE SIU LON

    RESISTÊNCIA ADESIVA AO CISALHAMENTO DE BRACKETS CERÂMICOS MONOCRISTALINOS: ESTUDO IN VITRO

    Dissertação apresentada ao Programa de Pós-

    Graduação em Odontologia da Pontifícia

    Universidade Católica do Paraná, como parte dos

    requisitos para obtenção do título de Mestre em

    Odontologia – Área de Concentração em

    Ortodontia.

    Orientador: Prof. Dr. Odilon Guariza Filho

    CURITIBA

    2008

  • Dados da Catalogação na Publicação

    Pontifícia Universidade Católica do Paraná

    Sistema Integrado de Bibliotecas – SIBI/PUCPR

    Biblioteca Central

    Lon, Luís Filipe Siu L847r Resistência adesiva ao cisalhamento de brackets cerâmicos 2008 monocristalinos : estudo in vitro / Luís Filipe Siu Lon ; orientador, Odilon Guariza Filho. – 2008. 60 f. ; 30 cm Dissertação (mestrado) – Pontifícia Universidade Católica do Paraná, Curitiba, 2008 Inclui bibliografia 1. Ortodontia. 2. Braquetes ortodônticos. 3. Maloclusão. 4. Ortodontia corretiva. 5. Cerâmica odontológica. 6. Aparelhos ortodônticos. I. Guariza Filho, Odilon. II. Pontifícia Universidade Católica do Paraná. Programa de Pós-Graduação em Odontologia. III. Título. CDD 20. ed. – 617.643

  • DEDICATÓRIA À minha querida esposa Beatriz,

    pelo amor, carinho, companheirismo, paciência, dedicação e compreensão

    nas horas que estive ausente pela busca de um sonho;

    Aos meus filhos Luís Miguel e Luís Eduardo,

    a felicidade que transformou a minha vida. Ainda que privados da minha

    companhia devido à dedicação ao mestrado em Ortodontia, me ensinam muito a

    cada dia;

    Aos meus pais Lina e Yin,

    responsáveis pela minha formação intelectual e moral. A eles a minha eterna

    gratidão, pela dedicação e apoio durante a minha vida acadêmica;

    Aos meus irmãos Sérgio e Aurélio,

    pelo apoio, incentivo e amizade constante;

    Aos meus sogros Nelly e Young,

    pelo apoio e pelo exemplo de caráter para a formação moral dos meus filhos.

    DEDICO

  • AGRADECIMENTO ESPECIAL

    À Deus, pela dádiva da vida, força, coragem para perseguir meus objetivos e

    por sempre iluminar o meu caminho.

    Ao Prof. Dr. Odilon Guariza Filho, meu orientador, pela amizade construída

    ao longo desta jornada, pelos momentos bem humorados, confiança, sinceridade e

    orientação segura deste trabalho.

    Ao Prof. Dr. Orlando Tanaka, responsável pela área de concentração em

    Ortodontia da PUCPR, pelo exemplo, dedicação e competência junto ao curso. Serei

    sempre grato pela amizade, pela confiança em mim depositada e pelas valiosas

    contribuições que certamente me tornaram um profissional melhor preparado e uma

    pessoa mais experiente.

  • AGRADECIMENTOS

    À Pontifícia Universidade Católica do Paraná pela oportunidade de realizar

    este mestrado e pela bolsa de estudos parcial concedida.

    Ao Diretor do Programa de Pós-Graduação em Odontologia, Prof. Dr. Sérgio

    Vieira, pelo apoio e atenção dispensados.

    Ao Prof. Dr. Hiroshi Maruo, pelo convívio, amizade, pelas histórias de vida, e

    ensinamentos transmitidos.

    À Profa. Dra. Elisa Souza Camargo, pelo agradável convívio, conhecimentos

    transmitidos, dedicação e responsabilidade.

    Ao Prof. Dr. Júlio Wilson Vigorito, responsável pela minha formação

    ortodôntica, pelos seus ensinamentos e estímulo para a busca constante do

    conhecimento.

    Aos meus colegas da 6ª Turma de Mestrado: Bruno, Fernanda, Jucienne,

    Luciana, Luégya, Marcos, Mariana, Raul, Ricardo, Saulo e Taís, pela união durante

    os dois anos, que selou a nossa amizade.

    Aos alunos da 1a turma de Doutorado em Odontologia, área de concentração

    em Ortodontia da PUCPR, Armando e Ivan, pelo convívio agradável.

    Ao amigo Dr. Ivan Taffarel, pelo apoio, amizade e estímulo constante.

    Ao Prof. José Carlos Munhoz da Cunha, pela carta de recomendação.

    Ao Prof. Dr. Sérgio Aparecido Ignácio, pelo convívio, e competência na

    análise estatística.

    A todos os professores das áreas conexas, pelos ensinamentos transmitidos.

  • Ao técnico do laboratório de Ensaios de Materiais da Engenharia Mecânica,

    Jaison Sanders, pela disponibilidade e presteza durante a realização dos ensaios

    mecânicos.

    À Neide Borges dos Reis, secretária do Programa de Pós-Graduação da

    PUCPR, pelo carinho, e atenção prestada.

    Às estagiárias Aline Wiens e Paula Maiza de Marchi, pela atenção e serviços

    prestados.

    À Silvana Casagrande Gabardo pela presteza e auxílio durante a clínica.

    Aos funcionários da PUCPR, pela disponibilidade dos seus serviços.

    A todos que direta ou indiretamente contribuíram na elaboração deste

    trabalho.

    Aos pacientes do meu consultório que souberam ou não entender a minha

    ausência.

  • SUMÁRIO

    1 ARTIGO EM PORTUGUÊS 01

    PÁGINA TÍTULO 02

    RESUMO 03

    INTRODUÇÃO 04

    MATERIAL E MÉTODOS 06

    RESULTADOS 09

    DISCUSSÃO 11

    CONCLUSÕES 14

    REFERÊNCIAS 15

    TABELAS 19

    FIGURAS 20

    2 ARTIGO EM INGLÊS 22

    TITLE PAGE 23

    ABSTRACT 24

    INTRODUCTION 25

    MATERIAL AND METHODS 27

    RESULTS 30

    DISCUSSION 31

    CONCLUSIONS 34

    REFERENCES 35

    TABLES 39

    FIGURES 40

    3 ANEXOS 42

    ANEXO 1 – PARECER DO COMITÊ DE ÉTICA 43

    ANEXO 2 – MATERIAL E MÉTODOS 45

    ANEXO 3 – ANÁLISE ESTATÍSTICA 47

    ANEXO 4 – NORMAS PARA PUBLICAÇÃO – AMERICAN JOURNAL OF

    ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS 49

  • 1 ARTIGO EM PORTUGUÊS

  • 2

    PÁGINA TÍTULO

    RESISTÊNCIA ADESIVA AO CISALHAMENTO DE BRACKETS CERÂMICOS MONOCRISTALINOS: ESTUDO IN VITRO Luís Filipe Siu Lon Mestrando em Odontologia - Área de Concentração em Ortodontia Pontifícia Universidade Católica do Paraná (PUCPR) Curitiba, Paraná, Brasil Odilon Guariza Filho Professor adjunto do Programa de Pós-Graduação em Odontologia - Área de Concentração Ortodontia Pontifícia Universidade Católica do Paraná (PUCPR) Curitiba, Paraná, Brasil Endereço para correspondência: Prof. Dr. Odilon Guariza Filho Pontifícia Universidade Católica do Paraná Programa de Pós-Graduação em Odontologia - Ortodontia Rua Imaculada Conceição, 1155 Fone: 55 41 3271-1637 / Fax: 55 41 3271-1405 80.215-901 – Curitiba-PR – Brasil e-mail: [email protected]

  • 3

    Resistência adesiva ao cisalhamento de brackets cerâmicos monocristalinos:

    estudo in vitro

    RESUMO

    Introdução: A crescente procura por tratamentos ortodônticos pelos adultos em busca de correção da oclusão e estética, tem aumentado a demanda no uso de brackets cerâmicos. O objetivo deste trabalho foi avaliar, in vitro, a Resistência Adesiva (RA) ao cisalhamento e o modo da falha adesiva pelo Índice de Adesivo Remanescente (IAR) de brackets cerâmicos monocristalinos, Radiance® e Inspire Ice®, comparando-os com brackets metálicos Miniature Twin® e brackets cerâmicos policristalinos Clarity®. Métodos: Foram utilizados 80 incisivos inferiores permanentes bovinos, distribuídos em 4 grupos de 20 dentes cada: G I- brackets metálicos Miniature Twin® (3M Unitek®), G II- brackets cerâmicos policristalinos Clarity® (3M Unitek®), G III- brackets cerâmicos monocristalinos Radiance® (American Orthodontics®) e G IV- brackets cerâmicos monocristalinos Inspire Ice® (ORMCO®). Os brackets foram colados com Transbond™ XT (3M Unitek®). Após 24 h, avaliou-se a RA em máquina universal de ensaios e o IAR. Resultados: A RA média em MPa foi: G I= 13,23, G II= 21,00, G III= 12,08 e G IV=16,09. A análise de variância e o teste de Games-Howell indicaram que não houve diferença significante (P>0,05) entre o G I e G III e entre o G I e G IV. O G II diferiu estatisticamente (P

  • 4

    INTRODUÇÃO

    A partir da década de 60, surgiram os primeiros brackets não metálicos,

    fabricados em policarbonato, que apresentavam problemas quanto à embebição de

    água, distorções sob forças ortodônticas, descoloração e fratura.1-3

    Diante do insucesso inicial dos brackets de policarbonato, os brackets

    cerâmicos foram introduzidos na Ortodontia no final da década 80, visando a

    eliminação do problema estético dos brackets metálicos, sem a desvantagem dos

    brackets de policarbonato.3-5 Desde então, os fabricantes desenvolvem novas

    tecnologias no aperfeiçoamento dos brackets estéticos.

    Os brackets cerâmicos são produzidos a partir de óxido de alumínio (Al2O3) e

    podem ser policristalinos ou monocristalinos, estes últimos também conhecidos

    como brackets de safira devido a sua estrutura cristalina ser semelhante à estrutura

    da safira natural.6

    Os brackets policristalinos são produzidos pela precipitação de partículas de

    óxido de alumínio, combinadas a um aglutinador, trabalhadas e aquecidas para

    remover as imperfeições da superfície e o estresse causado pelo processo de corte.

    Este processo de fabricação pode ocasionar falhas estruturais no acessório.6

    Os brackets cerâmicos monocristalinos são fabricados pela fundição de

    partículas puras de óxido de alumínio a 2100°C, formando um único cristal de óxido

    de alumínio que resultará na fabricação do bracket. Esta massa é resfriada

    vagarosamente para permitir um cuidadoso controle da cristalização. Esse material é

    recortado no tamanho dos brackets, e técnicas de corte ultrasônico, diamantado ou

    a combinação de ambas são utilizadas para sua formatação.6,7 Desta maneira, a

    estrutura monocristalina dos brackets é mais pura, com menor número de falhas

    estruturais, o que torna a sua produção com um custo mais elevado que a dos

  • 5

    policristalinos. Estes brackets monocristalinos são mais estéticos, mais translúcidos

    e possuem a resistência maior à tração em comparação aos policristalinos.6-8

    Os primeiros brackets cerâmicos utilizavam o silano como agente de união

    para aumentar as pontes moleculares de união entre a base do bracket e o adesivo

    e, durante a remoção, poderiam resultar em fraturas do esmalte dentário.6,9,10

    As retenções mecânicas, edentações ou micro-retenções na base do bracket

    foram introduzidas para substituir a retenção química e reduzir a resistência adesiva

    dos brackets cerâmicos, evitando maiores danos ao esmalte dentário.5 Aprimorar e

    desenvolver novos brackets cerâmicos, mais estéticos, com resistência adesiva

    adequada, resistência à fratura, menor atrito e mínimo dano ao esmalte constitui

    uma necessidade para a melhoria do tratamento ortodôntico com estes brackets.11

    Dentre estes aprimoramentos estão os brackets cerâmicos monocristalinos Inspire

    Ice® (ORMCO®, Orange, CA) e o Radiance® (American Orthodontics®, Sheboygan,

    WI).

    Os brackets Inspire Ice® utilizam micro-esferas de zircônia fundidas na base

    para prover retenção mecânica. Entretanto, possui uma faixa de 0,021”, na região

    gengival, que não apresenta estas esferas para possibilitar um ponto fraco iniciador

    da falha adesiva durante a sua remoção (ORMCO®, Orange, CA). Os brackets

    Radiance® apresentam retenções micro-mecânicas no centro e ausência de

    retenções na periferia da base para facilitar a remoção do bracket (American

    Orthodontics®, Sheboygan, WI).

    A crescente procura por tratamentos ortodônticos pelos adultos em busca de

    correção da oclusão e estética, tem aumentado a demanda no uso de brackets

    cerâmicos.

  • 6

    Portanto, o objetivo deste trabalho foi avaliar, in vitro, a Resistência Adesiva

    (RA) ao cisalhamento e o modo da falha adesiva pelo Índice de Adesivo

    Remanescente (IAR) dos brackets cerâmicos monocristalinos, Radiance® e Inspire

    Ice®, comparando-os com brackets metálicos Miniature Twin® e brackets cerâmicos

    policristalinos Clarity®.

    MATERIAL E MÉTODOS

    O presente estudo foi aprovado com o parecer no 204/07 do Comitê de Ética

    no Uso de Animais (CEUA) da Pontifícia Universidade Católica do Paraná (PUCPR).

    A amostra consistiu de 80 incisivos inferiores permanentes bovinos, recém

    extraídos, com as coroas intactas, ausência de descalcificações, trincas e fraturas.

    Os dentes foram armazenados para desinfecção em Cloramina-T 0,5%, durante 7

    dias. Após este período, os tecidos moles foram removidos com o auxílio de curetas

    periodontais (Duflex, Juiz de Fora, Brasil), as raízes dentárias foram seccionadas no

    terço médio e a polpa dentária foi removida com lima endodôntica Hedstron

    (Maillefer, Ballaigues, Suíça), e depois os dentes foram armazenados em água

    destilada, a 4ºC, trocada semanalmente, conforme ISO-TR 11405.12

    Os dentes foram divididos aleatoriamente em 4 grupos de 20 dentes cada, e

    os brackets foram colados nas coroas dentárias, assim divididos (Fig. 1, página 20):

    Grupo I (controle): Brackets metálicos Miniature Twin®, slot retangular 0,022”,

    prescrição de Roth (017-313, 3M Unitek®, Monróvia, CA) para incisivos

    inferiores, com área da base de 7,23 mm2.

    Grupo II: Brackets cerâmicos policristalinos Clarity®, slot retangular 0,022”,

    prescrição de Roth (6400-721, 3M Unitek® non-coated, Monróvia, CA)

    para incisivos inferiores, com área da base de 8,64 mm2.

  • 7

    Grupo III: Brackets cerâmicos monocristalinos Radiance®, slot retangular 0,022”,

    prescrição de Roth (002-7142, American Orthodontics®, Sheboygan, WI)

    para incisivos inferiores, com área da base de 10,63 mm2.

    Grupo IV: Brackets cerâmicos monocristalinos Inspire Ice®, slot retangular 0,022”,

    prescrição de Roth (444-0011, ORMCO®, Orange, CA) para incisivos

    inferiores, com área da base de 11,49 mm2.

    Preparo dos corpos de prova

    Os dentes foram inicialmente posicionados numa morsa para os

    procedimentos de colagem. Foi realizada profilaxia com taça de borracha e pasta de

    pedra pomes (SS White, Rio de Janeiro, Brasil) com micromotor em baixa rotação,

    durante 10 segundos, na região da futura colagem (superfície mais plana do terço

    incisal vestibular da coroa) (Anexo 2, Fig. 1A e 1B, página 45). Utilizou-se uma taça

    de borracha para cada 5 dentes.

    Em seguida, os dentes foram lavados com jatos de água destilada da seringa

    tríplice por 20 segundos e secos com ar comprimido isento de umidade ou óleo por

    20 segundos. Para o condicionamento do esmalte, foi utilizado o ácido fosfórico a

    37% (Dentsply, Petrópolis, Brasil) por 30 segundos (Anexo 2, Fig. 1C, página 45). Os

    esmaltes condicionados foram lavados com água destilada por 20 segundos e secos

    por 20 segundos com ar comprimido isento de umidade ou óleo.

    Para a colagem dos brackets foi utilizado o adesivo fotoativado Transbond™

    XT (3M Unitek®, Monrovia, CA) seguindo-se as instruções do fabricante (Anexo 2,

    Fig. 2, página 45).

    Os brackets foram posicionados no terço incisal da superfície vestibular da

    coroa, com o auxílio de pinça para colagem (Morelli®, Sorocaba, Brasil), com o slot

    paralelo ao longo eixo do dente, para minimizar a possibilidade de fraturas das

  • 8

    aletas dos brackets cerâmicos. Uma carga de 400 gf de pressão foi aplicada com

    dinamômetro (ETM®, Monróvia, CA) para padronizar a espessura da resina e o

    excesso de resina escoada foi removido com sonda exploradora (Anexo 2, Fig. 1E,

    página 45).

    Foi realizada a fotopolimerização com o aparelho fotopolimerizador, de luz

    halógena, Optilux®, 500 (Demetron Kerr®, Danbury, CT) com irradiância de 600

    mW/cm2 por 20 segundos nos brackets metálicos, sendo 10 segundos na interface

    mesial e 10 segundos na interface distal (Anexo 2, Fig. 1F, página 46). Os brackets

    cerâmicos foram fotopolimerizados por 10 segundos com o feixe de luz incidindo

    perpendicularmente através dos brackets cerâmicos. Em todos os dentes a distância

    entre o feixe de luz e o bracket foi de 5 mm, conforme instruções do fabricante do

    Transbond™ XT.

    Utilizou-se um dispositivo de aço inoxidável com um fio retangular 0,021” x

    0,025” (Unitek®, Monróvia, CA) soldado 90º em relação à sua base, para garantir a

    perpendicularidade do bracket em relação ao plano horizontal. Os dentes foram

    então fixados no dispositivo por uma ligadura elástica (Tp Orthodontics, La Port, IN)

    colocada nas aletas do bracket, prendendo-o a este fio retangular (Figuras 2A, 2B e

    2C, página 20). Após a fixação, a raiz dentária foi envolvida por um anel metálico

    (2,5 cm de diâmetro e 2,0 cm de altura), e foi preenchido com resina acrílica

    autopolimerizável (Jet®, São Paulo, Brasil) (Fig. 2D, página 20). Desta forma a

    superfície de colagem ficou perpendicular ao plano horizontal e paralelo à direção de

    aplicação da força no ensaio de cisalhamento (Fig. 2E, página 20). Os corpos de

    prova ficaram armazenados em água destilada à temperatura de 37ºC por 24 horas.

    Os corpos de prova foram submetidos ao teste de resistência adesiva ao

    cisalhamento em uma máquina universal de ensaios EMIC® DL 500 (Emic

  • 9

    Equipamentos e Sistemas de Ensaio Ltda., São José dos Pinhais, Brasil), a uma

    velocidade de 0,5 mm/minuto e com célula de carga de 50 kgf no sentido ocluso-

    gengival. Um computador conectado à máquina de ensaios registrou a tensão de

    ruptura em MPa, de cada corpo de prova considerando a área da base de cada

    bracket (Fig. 2F, página 20).

    Após a descolagem dos brackets, a superfície do esmalte e os brackets foram

    examinados com uma lupa estereoscópica Leica Zoom 2000 (Leica®, Wetzlar,

    Alemanha) com 20 vezes de aumento, para a verificação do IAR. O adesivo

    remanescente foi graduado de acordo com o método utilizado por Årtun e

    Bergland13, variando de 0 a 3. A escore (0) indica que não há resina aderida à

    superfície de esmalte do dente; (1) indica que menos da metade da resina ficou

    aderida à superfície de esmalte do dente; (2) indica que mais da metade da resina

    ficou aderida à superfície de esmalte do dente; (3) indica que toda a resina ficou

    aderida, inclusive a impressão da malha do bracket.

    Espécimes representativos do IAR de cada grupo, foram visualizados sob

    Microscopia Eletrônica de Varredura (MEV) (SSX 550, SuperScan, Shimadzu, Kyoto,

    Japão) (Fig. 3, página 21).

    A análise estatística foi feita com o software Statistical Package for Social

    Science 13.0 for Windows (SPSS, Inc., Chicago, IL) e utilizou-se os seguintes testes

    estatísticos: Kolmogorov-Smirnov, Levene, ANOVA a um critério, Games-Howell e

    de Kruskal-Wallis.

    RESULTADOS

    Os resultados da resistência adesiva estão descritos na Tabela I (página 19).

  • 10

    Inicialmente testou-se a normalidade dos dados pelo teste de Kolmogorov-

    Smirnov ao nível de significância de 0,05 o qual mostrou que apenas o grupo I

    (Miniature Twin®) não apresentou normalidade, uma vez que p=0,04305 (Anexo 3,

    Tabela I, página 47). Testou-se ainda, a homogeneidade de variância por meio do

    teste de Levene, evidenciando heterogeneidade entre os grupos (p

  • 11

    significantes para o IAR, uma vez que p>0,05 (Tabela II, página 20 e Anexo 3,

    Figura 2, página 48).

    DISCUSSÃO

    O presente estudo avaliou a Resistência Adesiva (RA) ao cisalhamento dos

    brackets cerâmicos monocristalinos (Radiance® e Inspire Ice®) e o modo de falha

    adesiva dos mesmos comparando-os aos brackets metálicos e cerâmicos

    policristalinos, cuja RA já fora amplamente estudada e apresentada na

    literatura.1,2,5,9,10 Para tanto, foram utilizados incisivos inferiores permanentes

    bovinos, cujo o esmalte é semelhante ao humano em suas propriedades físicas e

    composições.14,15 No entanto, Oesterle et al.16 relataram que, no esmalte bovino, a

    RA é 21% a 44% menor em relação ao esmalte humano, devido a uma formação

    mais rápida do esmalte e dentina durante o desenvolvimento dental do dente bovino.

    A colagem de brackets cerâmicos ao esmalte pode ocorrer pelos seguintes

    processos: (1) adesão química por meio de agente silanizador para união da base

    com o adesivo, (2) retenção mecânica via edentações na base e (3) combinação das

    duas.6,9

    A adesão química pode promover maior RA quando comparada com a

    retenção mecânica,9,10,17-19 entretanto, outras pesquisas mostraram que não houve

    aumento na RA utilizando silano como agente de união. 20-23

    Um sistema de adesão aceitável em Ortodontia deve resistir às forças

    ortodônticas e mastigatórias e permitir fácil remoção dos brackets, com mínimo ou

    nenhum dano ao esmalte.24,25

    No presente estudo, os valores médios de RA foram de 13,23 ± 5,15 MPa,

    21,00 ± 5,09 MPa, 12,08 ± 2,13 MPa e 16,09 ± 4,16 MPa, respectivamente para

  • 12

    Miniature Twin®, Clarity®, Radiance® e Inspire Ice®, valores acima da faixa

    clinicamente aceitável de acordo com Reynolds.26 Este autor sugeriu que os valores

    de RA entre 5,9 a 7,8 MPa seriam adequados para a maioria das necessidades

    clínicas. Portanto, todas as marcas estudadas apresentaram adesão adequada para

    uso clínico.

    Alguns trabalhos indicam que os brackets cerâmicos têm a RA maior que os

    brackets metálicos, o que concorda com os resultados deste estudo.9,27-29 A média

    da RA, obtida nesta pesquisa, para os brackets Clarity® (21,00 ± 5,09 MPa) foi

    semelhante aos resultados de Theodorakopoulou,30 porém maiores que os

    encontrados em outros estudos.31-33 Bishara et al.31 testaram a RA dos brackets

    cerâmicos Clarity® colados em esmalte com Transbond™ XT e obtiveram um valor

    médio de 10,4 ± 4,1 MPa. Mundstock et al.33 encontraram valores médios de RA de

    13,27 ± 5,39 MPa. Estes trabalhos realizaram os testes de cisalhamento entre 5 e 30

    minutos após os procedimentos de colagem, período no qual ainda não ocorreu

    completamente a polimerização da resina composta. No presente estudo os testes

    de cisalhamento foram realizados 24 horas após a colagem e estas diferenças,

    provavelmente, podem ser devido ao tempo de espera para a realização dos testes

    de cisalhamento, pois conforme demonstrado por Yamamoto et al.34 e Oesterle e

    Shellhart 35 a RA aumenta progressivamente nas primeiras 24 horas após a

    polimerização do bracket. Entretanto, Liu et al.36 realizaram os testes de RA 24

    horas após a colagem e encontraram uma média de 11,83 ± 3,46 MPa para o

    Clarity®.

    A RA dos brackets Radiance® foi de 12,08 ± 2,13 MPa, valores inferiores em

    relação aos outros brackets testados neste estudo, porém, acima da faixa de valores

    preconizados por Reynolds26. Estes valores podem ser devido à diminuição da área

  • 13

    de retenção da base do bracket, incorporado pelo fabricante, com o intuito de

    facilitar a remoção do bracket sem dano ao esmalte dentário, pois de acordo com

    Joseph e Rossow,27 Ghafari et al.,37 e Sinha et al.,38 o tipo de retenção e a base do

    bracket, são fatores que influenciam a RA. Até o momento, não há estudos

    envolvendo a RA dos brackets Radiance®.

    Os brackets Inspire Ice® apresentaram valores médios de RA 16,09 ± 4,16

    MPa, resultados similares de Theodorakopoulou et al.30 e próximos de Liu et al.36,

    que obtiveram valores médios de 20,32 ± 8,00 MPa e 11,95 ± 4,60 MPa

    respectivamente utilizando brackets Inspire®. Entretanto, vale ressaltar que os

    brackets Inspire Ice® são uma evolução do Inspire, com redução da área de retenção

    visando a diminuição da RA. Isto porque, forças de adesão superiores a 13,73 MPa

    deveriam ser evitadas para eliminar a possibilidade de fraturas de esmalte durante a

    descolagem dos brackets.39,40

    Para manter a integridade do esmalte dentário a falha adesiva ideal deve

    ocorrer na interface bracket-adesivo.38 Os escores do IAR idealizados por Årtun e

    Bergland13 têm a finalidade de quantificar o material remanescente sobre o esmalte,

    mostrando o local onde ocorreu a falha no teste de cisalhamento.

    Nos brackets metálicos Miniature Twin® o escore do IAR predominante foi o 1,

    com uma freqüência de 70%, evidenciando falha na interface adesivo-esmalte, o que

    aumenta as possibilidades de trincas no esmalte durante a sua remoção (Tabela II,

    página 19). Nos brackets Clarity®, Radiance® e Inspire Ice® a falha adesiva ocorreu

    na interface bracket-adesivo com grande maioria apresentando escore 3 de IAR, ou

    seja, todo o adesivo ficou aderido ao dente, com freqüência de 80%, 75% e 65%

    respectivamente (Tabela II, página 19). Estes resultados estão em concordância

    com outros estudos que utilizaram brackets cerâmicos, os quais apresentam como

  • 14

    vantagem a manutenção da integridade do esmalte, e como desvantagem a

    necessidade de remoção mecânica do adesivo remanescente.30-33

    As diferenças numéricas no resultado da RA dos brackets deste estudo,

    quando comparado aos de outros,31-33 provavelmente, deve-se a distintas

    metodologias e substratos utilizados. Este estudo demonstrou que apesar da RA dos

    brackets cerâmicos ser mais alta que os metálicos, a falha adesiva ocorreu na

    interface bracket-adesivo, minimizando as chances de danos ao esmalte. Isto torna

    os brackets cerâmicos uma alternativa viável e segura para o uso clínico.

    CONCLUSÕES

    • Os brackets Radiance® e Inspire Ice® apresentaram RA semelhante aos

    brackets metálicos;

    • Os brackets Clarity® apresentaram RA maior que os dois brackets

    monocristalinos estudados;

    • Os brackets cerâmicos monocristalinos mostraram padrão de falha na interface

    bracket-adesivo, sugerindo a tendência do adesivo ficar aderido ao dente após

    a sua remoção.

  • 15

    REFERÊNCIAS

    1. Chaconas S, Caputo A, Niu G. Bond strength of ceramic brackets with various

    bonding systems. Angle Orthod 1990; 61:35-42.

    2. Britton JC, McInnes P, Weinberg R, Ledoux WR, Retief DH. Shear bond strength

    of ceramic orthodontic brackets to enamel. Am J Orthod Dentofacial Orthop 1990;

    98:348-53.

    3. Karamouzos A, Athanasios E, Papadopoulos MA. Clinical characteristics and

    properties of ceramic brackets: A comprehensive review. Am J Orthod Dentofacial

    Orthop 1990; 112:34-40.

    4. Blalock KA, Powers JM. Retention capacity of the bracket bases of new esthetic

    orthodontic brackets. Am J Orthod Dentofacial Orthop 1995; 107:596-603.

    5. Bordeaux JM, Moore RN, Bagby MD. Comparative evaluation of ceramic bracket

    base designs. Am J Orthod Dentofacial Orthop 1994; 105:552-60.

    6. Swartz ML. Ceramic brackets. J Clin Orthod 1988; 22:82-8.

    7. Omana HM, Moore RN, Bagby MD. Frictional properties of metal and ceramic

    brackets. J Clin Orthod 1992; 26:425-432.

    8. Birnie D. Ceramic brackets. Br J Orthod 1990; 17:71-75.

    9. Bishara SE, Fehr DE, Jacobsen JR. A comparative study of the debonding

    strengths of different ceramic brackets, enamel conditioners, and adhesives. Am

    J Orthod Dentofacial Orthop 1993;104:170-9.

    10. Wang WN, Meng CL, Tarng TH. Bond strength: a comparison between chemical

    coated and mechanical interlock bases of ceramic and metal brackets. Am J

    Orthod Dentofacial Orthop 1997; 111:374-81.

    11. Bishara SE. Ceramic brackets and the need to develop national standards. Am J

    Orthod Dentofacial Orthop 2000; 117:595-597

  • 16

    12. International Organization for Standardization. Technical Specification. Dental

    Materials – Testing of adhesion to tooth structure. ISO/TS 11405, Second

    edition, 2003 (E).

    13. Årtun J, Bergland S. Clinical trials with crystal growth conditioning as an

    alternative to acid-etch enamel pretreatment. Am J Orthod 1984; 85:333-40.

    14. Smith H, Casko J, Leinfelder K, Utley JD. Comparison of orthodontic bracket

    bond strengths: human vs. bovine enamel. J Dent Res 1976; 55:153-67.

    15. Nakamichi I, Iwaku M, Fusayama T. Bovine teeth as possible substitutes in the

    adhesion test. J Dent Res 1983; 62:1076-81.

    16. Oesterle L, Shellhart W, Belanger G. The use of bovine enamel in bonding

    studies. Am J Orthod Dentofacial Orthop 1998; 114:514-9.

    17. Iwamoto H, Kawamoto T, Kinosita Z. Bond strength of new ceramic brackets as

    studied in vitro. J Dent Res 1987; 66:928.

    18. Viazis A D, Cavanaugh G, Bevis RR. Bond strengths of ceramic bracket under

    shear stress: An in vitro report. Am J Orthod Dentofacial Orthop 1990;98:214-21.

    19. Redd TB, Shivapuja PK. Debonding ceramic brackets: effects on enamel. J Clin

    Orthod 1991; 25:475-81.

    20. Guess MB, Watanabe LG, Beck FM, Crall MG. The effect of silane coupling

    agents on the bond strength of a polycrystalline ceramic bracket. J Clin Orthod

    1988; 22:788-92.

    21. Gwinnett AJ. A comparison of shear bond strengths of metal and ceramic

    brackets. Am J Orthod Dentofacial Orthop 1988; 93:346-8.

    22. Merrill SW, Oesterle LJ, Hermesch CB. Ceramic bracket bonding: a comparison

    of shear, tensile, and torsional bond strengths of ceramic brackets. Am J Orthod

    Dentofacial Orthop 1994;106:290-7.

  • 17

    23. Ødegaard J, Segner D. Shear bond strength of metal brackets compared with a

    new ceramic bracket. Am J Orthod Dentofacial Orthop 1988; 94:201-6.

    24. Zachrisson BU. Bonding in Orthodontics. In: Graber TM, Swain BF, editors.

    Orthodontic, current principles and techniques. St Louis: Mosby; 2005.

    25. Chay SH, Wong SL, Mohamed N, Chia A, Yap AU. Effects of surface treatment

    and aging on the bond strength of orthodontic brackets to provisional materials.

    Am J Orthod Dentofacial Orthop 2007;132:577.e7-577.e11.

    26. Reynolds IR. A review of direct orthodontic bonding. Br J Orthod 1975; 2:171-8.

    27. Joseph VP, Rossouw E. The shear bond strengths of stainless steel and ceramic

    brackets used with chemically and lightactivated composite resins. Am J Orthod

    Dentofacial Orthop 1990; 97:121-5.

    28. Harris AMP, Joseph VP, Rossouw E. Comparison of shear bond strengths of

    orthodontic resins to ceramic and metal brackets. J Clin Orthod 1990; 24:725-8.

    29. Weinberger SJ, Foley TF, McConnell RJ, Wright GZ. Bond strengths of two

    ceramic brackets using argon laser, light, and chemically cured resin systems.

    Angle Orthod 1997; 3:173-8.

    30. Theodorakopoulou LP, Sadowsky PL, Jacobson A, Lacefield WJ. Evaluation of

    the debonding characteristics of 2 ceramic brackets: An in vitro study. Am J

    Orthod Dentofacial Orthop 2004; 125:329-36.

    31. Bishara SE, Olsen ME, Von Wald. Evaluation of debonding characteristics of a

    new collapsible ceramic bracket. Am J Orthod Dentofacial Orthop 1997; 112:

    552-9.

    32. Bishara SE, Olsen ME, Von Wald L, Jakobsen JR. Comparison of the debonding

    characteristics of two innovative ceramic bracket designs. Am J Orthod

    Dentofacial Orthop 1999; 116:86-92.

  • 18

    33. Mundstock KS, Sadowsky PL, Lacefield W, Bae S. An in vitro evaluation of a

    metal reinforced orthodontic ceramic bracket. Am J Orthod Dentofacial Orthop

    1999; 116:635-41.

    34. Yamamoto A, Yoshida T, Tsubota K, Takamizawa T, Kurokawa H, Miyazaki M.

    Orthodontic bracket bonding: Enamel Bond strength vs time. Am J Orthod

    Dentofacial Orthop 2006; 130:435.e1-435.e6.

    35. Oesterle LJ, Shellhart WC. Effect of aging on the shear bond strength of

    orthodontic brackets. Am J Orthod Dentofacial Orthop 2008;133:716-20.

    36. Liu JK, Chung CH, Chang CY, Shieh DB. Bond strength and debonding

    characteristics of a new ceramic bracket. Am J Orthod Dentofacial Orthop 2005;

    128:761-5.

    37. Ghafari J, Skanky T, Mante F. Shear bond strength of two ceramic brackets. J

    Clin Orthod 1992; 24: 491-3.

    38. Sinha PK, Nanda RS, Duncanson MG, Hosier MJ. Bond strengths and remnant

    adhesive resin on debonding for orthodontic bonding techniques. Am J Orthod

    Dentofacial Orthop 1995; 108:302-7.

    39. Retief DH. Failure at the dental adhesive etched enamel interface. J Oral Rehab

    1974; 1:265-84.

    40. Bishara SE, Fehr DE. Ceramic brackets: something old, something new, a

    review. Semin Orthod 1997; 3:178-88.

  • 19

    TABELAS

    Tabela I. Estatística descritiva da resistência adesiva (MPa), segundo grupos

    Grupos n Média Mediana Desvio Padrão Games-Howell

    Miniature Twin® 20 13,245 11,110 5,155 a, c Clarity® 20 21,006 21,101 5,090 b Radiance® 20 12,081 11,984 2,139 a Inspire Ice® 20 16,096 16,184 4,164 c Nota: Letras iguais significam que não há diferenças significantes (P

  • 20

    FIGURAS

    A1 B1

    Fig 1. Imagens dos brackets utilizados no estudo e suas respectivas bases visualizadas no MEV (24 X). A1; A2, Miniature Twin®; B1; B2, Clarity®; C1; C2, Radiance® e D1; D2, Inspire Ice®.

    D2

    A2 B2 C2

    Fig 2. Seqüência da montagem dos corpos de prova. A; B, dispositivo com o dente preso no fio 0,021” X 0,025” com ligadura elástica; C, anel metálico de inclusão posicionado para o preenchimento com acrílico autopolimerizável; D, preenchimento com acrílico; E, corpo de prova com o bracket perpendicular à base e paralelo à direção da força de cisalhamento e F, ensaio da RA ao cisalhamento.

    E F

    B A C

    D

    C1 D1

  • 21

    A D C B

    Fig 3. Imagens do MEV (24X) da superfície dentária, com o IAR representativo de cada grupo. A, Miniature Twin®- IAR 1; B, Clarity®- IAR 3; C, Radiance®- IAR 3 e D, Inspire Ice®- IAR 3.

  • 22

    2 ARTIGO EM INGLÊS

  • 23

    TITLE PAGE

    SHEAR BOND STRENGTH OF MONOCRYSTALLINE CERAMIC BRACKETS: IN

    VITRO STUDY

    Luís Filipe Siu Lon, DDS Graduate Dentistry Program – Orthodontics Master of Science Student Pontifical Catholic University of Paraná, Curitiba, Brazil Odilon Guariza Filho, DDS, MSD, PHD Associate Professor, Graduate Dentistry Program – Orthodontics Pontifical Catholic University of Paraná, Curitiba, Brazil Correspondence address: Odilon Guariza Filho, DDS, MSD, PHD Associate Professor, Graduate Dentistry Program – Orthodontics Pontifical Catholic University of Paraná Graduate Dentistry Program - Orthodontics Rua Imaculada Conceição, 1155 80215-901 Curitiba – Paraná – Brazil Phone: 55 – 41 - 3271-1637 Fax: 55 – 41 - 3271-1405 E-mail: [email protected]

  • 24

    Shear bond strength of monocrystalline ceramic brackets: in vitro study. ABSTRACT

    Introduction: The purpose of this study was to evaluate the Shear Bond Strength (SBS) and the adhesive failure mode of Radiance® (American Orthodontics, Sheboygan, Wisc) and Inspire Ice® (ORMCO, Orange, Calif) monocrystalline ceramic brackets and compare them with a metal bracket (Miniature Twin, 3M Unitek, Monrovia, Calif) and a polycrystalline ceramic bracket (Clarity, 3M Unitek, Monrovia, Calif). Methods: Eighty lower permanent bovine incisors were randomly divided into 4 groups of 20 teeth each: G I- Miniature Twin® metal brackets, G II- Clarity® polycrystalline ceramic brackets, G III- Radiance® monocrystalline ceramic brackets and G IV- Inspire Ice® monocrystalline ceramic brackets. The brackets were bonded with Transbond™ XT (3M Unitek) orthodontic adhesive. After 24 hours, the SBS were assessed on a universal testing machine and the Adhesive Remnant Index (ARI) was determined to show the adhesive failure mode. Results: The mean shear bond strength (MPa) was 13.23, 21.00, 12.08, and 16.09 for GI, GII, GIII, and GIV, respectively. Significant differences (p < 0.05) among groups were found, except between GI and GIII and GI and GIV. In most of the GI specimens, the adhesive failure occurred at the adhesive-enamel interface. For all the ceramic brackets tested, most of the specimens failed at the bracket-adhesive interface. Conclusions: The monocrystalline ceramic brackets presented an adequate SBS for clinical use and adhesive failure at bracket-adhesive interface, suggesting a tendency of the adhesive to remain on the tooth, reducing the enamel damage.

  • 25

    INTRODUCTION

    The first non-metallic brackets were introduced in orthodontics in the 1960s

    and were made of polycarbonate. These presented problems like water absorption,

    distortion under orthodontic forces, staining, and fracture.1-3 The first ceramic

    brackets were introduced in the late 80s to eliminate the aesthetic problem of metal

    brackets, without the disadvantages of polycarbonate brackets.3-5 Since then,

    manufacturers have developed new technologies to improve aesthetic brackets.

    Ceramic brackets are produced from aluminum oxide (Al2O3) and may be

    polycrystalline or monocrystalline, also known as sapphire brackets due to their

    crystalline structure, which resembles a natural sapphire.6

    Polycrystalline brackets are manufactured by blending aluminum oxide with a

    binder so that the mixture can be molded and burned to fuse alumina particles. The

    fused part is then machined to provide a bracket design and heated to remove

    surface imperfections and relieve stresses created by the cutting operations. This

    manufacturing process can cause structural failure in the bracket.6

    Monocrystalline ceramic brackets are produced by making a molten mass of

    pure aluminum oxide particles at temperatures in excess of 2100ºC, resulting in a

    single alumina crystal. The molten mass is slowly cooled to allow for controlled

    crystallization. The bracket is machined from the resulting crystal using ultrasonic

    cutting techniques, diamond cutting, or a combination of both.6,7 Therefore, the

    monocrystalline brackets are purer, with fewer imperfections and are more expensive

    than polycrystalline. The monocrystalline ceramic brackets are more aesthetically

    pleasing and translucent and had higher tensile strength than polycrystalline ones.6-8

  • 26

    A silane coupling agent was used in the early ceramic brackets as a chemical

    mediator at bracket base, increasing the molecular bridges between bracket base

    and the adhesive which could result in enamel cracking when debonding.6,9,10

    Mechanical retention with undercuts or micromechanical interlocking in

    ceramic bracket bases was introduced to substitute chemical retention, reducing the

    Shear Bond Strength (SBS) and decreasing the enamel damage.5 The development

    of ceramic brackets aesthetically pleasing with adequate SBS, fracture resistance,

    and low frictional resistance without enamel damage when debonding are

    characteristics of new monocrystalline ceramic brackets.11 Inspire Ice® (ORMCO®,

    Orange, CA) and Radiance® (American Orthodontics®, Sheboygan, WI)

    monocrystalline ceramic brackets seems to present these features.

    The Inspire Ice brackets use zirconia balls fused in the base to provide

    mechanical retention. However, there is a .021” strip in the gingival area that does

    not have these zirconia balls and reduces the force required for debonding, allowing

    a weaker point that starts the adhesive failure when debonding. The Radiance

    brackets base displays micromechanical interlock in the center and no retention on

    the edges for easy bracket debonding.

    An increasing number of adults seeking orthodontic treatment for occlusion

    correction and aesthetic purposes has created an increasing demand for ceramic

    brackets.

    In light of this, the objective of this study was to evaluate the Shear Bond

    Strength (SBS) and the failure mode using the Adhesive Remnant Index (ARI) of

    monocrystalline ceramic brackets, Radiance® and Inspire Ice® and to compare them

    with metal (Miniature Twin, 3M Unitek, Monrovia, CA) and polycrystalline ceramic

    brackets (Clarity, 3M Unitek, Monrovia, CA).

  • 27

    MATERIAL AND METHODS

    The present study was approved by the ethic committee of research of the

    Pontifical Catholic University of Paraná (PUCPR).

    Eighty freshly extracted bovine permanent incisors were used in the study.

    The criteria for tooth selection included the integrity of crowns, absence of decay,

    cracks, or fractures. The teeth were stored in a 0.5% Chloramine-T solution, for 7

    days for disinfection. After this period, soft tissues were removed using periodontal

    instruments (Duflex, Juiz de Fora, Brazil), roots were cut in the middle third, and the

    pulp was removed with Hedstron endodontic files (Maillefer, Ballaigues, Switzerland).

    The teeth were then stored in distilled water at 4ºC, which was replaced weekly

    according to ISO-TR 11405.12

    The teeth were randomly divided into 4 groups of 20 each (Fig.1):

    Group I (control): Miniature Twin metal brackets for lower incisor, with rectangular

    .022” slot (017-313, 3M Unitek®, Monróvia, CA). The average base area was

    7.23 mm2.

    Group II: Clarity® polycrystalline ceramic brackets for lower incisor, with rectangular

    .022” slot (6400-721, 3M Unitek® non-coated, Monróvia, CA). The average

    base area was 8.64 mm2.

    Group III: Radiance® monocrystalline ceramic brackets for lower incisor, with

    rectangular .022” slot (002-7142, American Orthodontics® Sheboygan, WI).

    The average base area was 10.63 mm2.

    Group IV: Inspire Ice® monocrystalline ceramic brackets for lower incisor, with

    rectangular .022” slot (444-0011, ORMCO®, Orange, CA). The average

    base area was 11.49 mm2.

  • 28

    Bonding Procedure

    Each tooth was cleaned, using a mixture of water and pumice, with a rubber

    polishing cup in the flattest buccal surface, using a low speed handpiece for 20

    seconds. The teeth were rinsed with distilled water and dried with oil-free

    compressed air for 20 seconds. A 37% phosphoric acid gel was applied to the buccal

    surface of each tooth for 30 seconds. The teeth were then rinsed with distilled water

    for 20 seconds and dried with oil-free compressed air for 20 seconds. Transbond™

    XT (3M Unitek®, Monrovia, CA) orthodontic light-cure composite was used. Bonding

    procedures were performed according to the manufacturer's instructions.

    The brackets were positioned in the flat buccal surface of the crown with the

    bracket slot parallel to the long-axis of the tooth to minimize “wing fracture” in the

    ceramic brackets. Each bracket was subjected to 400 gf using a dynamometer

    (ETM®, Monróvia, CA) to standardize the thickness of bonding resin. The excess of

    the resin was removed with a small scaller.

    The samples were light cured with Optilux®, 500 (Demetron Kerr®, Danbury,

    CT) at 600 mW/cm2. Each bracket was light-cured according to Transbond XT

    instructions. The light curing was performed for 20 seconds for metal brackets; 10

    seconds on the mesial interface, and 10 second on the distal interface of the bracket.

    The ceramic brackets were light-cured for 10 seconds through the brackets.

    A stainless steel device was made with a rectangular .021” X.025” wire (3M

    Unitek®, Monróvia, CA) welded in 90º in relation to the device base to ensure that the

    bracket was perpendicular in a horizontal plane. Then teeth were fixed on the device

    by an elastomeric ring holding it to the wire (Figs. 2A and B). After fixation, the root

    was embedded in chemically cured acrylic resin and placed in metal rings (Figs. 2C

    and D). That way, the bonding surface was perpendicular to the horizontal plane end

  • 29

    parallel to the direction of the force to be applied in shear bond strength test (Fig

    .2E). All bonded specimens were stored in distilled water at 37º C.

    Twenty-four hours after the bonding procedures, the shear bond strength

    (SBS) test was performed on an EMIC DL500 universal testing machine (Emic

    Equipamentos e Sistemas de Ensaio Ltda., São José dos Pinhais, Brazil) (Fig. 2F),

    with a 50 kgf load cell, at 0.5 mm/min crosshead speed, until the bracket ruptured. A

    computer connected to the machine registered the result in MPa of each tooth

    submitted to the test, considering the bracket area.

    Subsequently, the teeth were examined with a stereoscopic binocular Leica

    Zoom 2000 ( Leica®, Wetzlar, Germany) under 20X magnification to verify the

    Adhesive Remnant Index (ARI), according Årtun and Bergland13 scored from 0 to 3.

    The ARI scoring scale was: (0) no adhesive remained on the enamel, (1) less than

    50% of the adhesive remained on the enamel, (2) more than 50% of the adhesive

    remained on the enamel, and (3) all adhesive remained on the enamel, including the

    bracket mesh impression. Representative IAR specimens of each group were then

    chosen to be analyzed by Scanning Electronic Microscopy (SEM) (SSX 550,

    SuperScan, Shimadzu, Kyoto, Japan) (Fig.3).

    The results were analyzed by Statistical Package for Social Science 13.0 for

    Windows software (SPSS, Inc., Chicago, IL). Kolmogorov-Smirnov and Levene tests

    evaluated treatment normality and homogeneity. The results of the shear bond

    strengths were analyzed by one-way analysis of variance (ANOVA) and Games-

    Howell. The Kruskal-Wallis multiple comparisons test was used to analyze ARI

    scores.

  • 30

    RESULTS

    The descriptive statistics for the shear bond strength in megapascals (MPa)

    are shown in Table I for all groups. To verify normality, the Kolmogorov-Smirnov test

    was performed and only group I (Miniature Twin®) did not present normality at a 0.05

    level of significance. The Levene test demonstrated heterogeneity of variance among

    groups.

    When one-way ANOVA demonstrated a statistically significant difference

    among groups, the Games-Howell test was used with a level of significance at 0.05.

    There was a statistically significant difference between groups I (Miniature Twin®)

    and II (Clarity®). There was no statistically significant difference (P < 0.05) between

    groups I (Miniature Twin®) and III (Radiance®), or groups I (Miniature Twin®) and IV

    (Inspire Ice®). The shear bond strength of group II (Clarity®) was statistically

    significant (P < 0.05) than groups I, III, or IV. Statistically significant (P < 0.05)

    differences were found between groups III (Radiance®) and IV (Inspire Ice®)

    The relative frequencies of ARI scores for the four groups are presented in

    Table II. Kruskal-Wallis indicated that there was a statistically significant difference (P

    > 0.05) for ARI in group I compared to all other groups. Groups II, III, and IV were not

    statistically different.

    Adhesive failure occurred in the adhesive-enamel interface in 70% of the

    group I specimens (Table II). For groups II, III, and IV, the bond failure mode

    occurred at bracket-adhesive interface in 80%, 75%, and 65% respectively (Table II).

  • 31

    DISCUSSION

    The present study assessed the Shear Bond Strength (SBS) of

    monocrystalline ceramic brackets (Radiance® and Inspire Ice®) and their failure mode

    in comparison with metal and polycrystalline brackets, whose SBS has been

    extensively studied and presented in the literature.1,2,5,9,10 Bovine mandibular

    permanent incisors were used to assess SBS because bovine and human enamel

    have similar physical properties and compositions and can be successfully used for

    bonding tests to enamel.14,15 However, Oesterle et al.16 reported that the SBS of

    bovine enamel is between 21% and 44% lower than that of human enamel due to

    faster formation of enamel and dentin during the dental development of the bovine

    tooth.

    Bonding of ceramic brackets to enamel can occur by the following processes:

    (1) chemical bonding by means of a silanizing agent for bonding the base with

    adhesive, (2) mechanical retention via indentations in the base, and (3) a

    combination of the two.6,9

    Chemical bonding can promote higher SBS when compared with mechanical

    retention,9,10,17-19 however, the results of other studies have shown that there was no

    increase in SBS when silane was used as a bonding agent. 20-23

    An acceptable bonding system in Orthodontics must resist orthodontic and

    masticatory forces and allow for easy removal of brackets with minimum or no

    damage to enamel.24,25 Reynolds26 suggested that SBS values between 5.9 and 7.8

    MPa would be adequate for the majority of clinical requirements.

    In the present study, the mean SBS values were 13.23 ± 5.15 MPa, 21.00 ±

    5.09 MPa, 12.08 ± 2.13 MPa and 16.09 ± 4.16 MPa for Miniature Twin®, Clarity®,

    Radiance® and Inspire Ice®, respectively, which are all above the clinically acceptable

  • 32

    range suggested by Reynolds.26 Therefore, all the brands studies presented

    adequate bonding for clinical use.

    Some studies indicate that ceramic brackets have higher SBS values than

    metal brackets, which is in agreement with the results of this study.9,27-29 The mean

    SBS obtained in this research for Clarity® brackets (21.00 ± 5.09 MPa) was similar to

    the results of Theodorakopoulou,30 but higher than those found in other studies that

    used these brackets.31-33 Bishara et al.31 tested the SBS of Clarity® ceramic brackets

    bonded to enamel with Transbond™ XT and obtained a mean value of 10.4 ± 4.1

    MPa. Mundstock et al.33 found mean SBS values of 13.27 ± 5.39 MPa. These studies

    performed shear tests between 5 and 30 minutes after the bonding procedures, a

    period in which complete polymerization of resin composite had not yet occurred. In

    the present study, the shear tests were performed 24 hours after bonding and these

    differences are probably due to the time that elapsed before the shear tests were

    performed, as shown by Yamamoto et al.34 and Oesterle and Shellhart.35

    Nevertheless, Liu et al.36 performed SBS tests 24 hours after bonding and found a

    mean of 11.83 ± 3.46 MPa for Clarity®.

    The shear bond strength of Radiance® brackets was 12.08 ± 2.13 MPa, lower

    in comparison with the other brackets tested in this study but higher than the range of

    values recommended by Reynolds26. These values can be due to diminishment of

    the bracket base retention area incorporated by the manufacturer, with the purpose

    of making easier the bracket removal without damaging dental enamel. According to

    Joseph and Rossow,27 Ghafari et al.,37 and Sinha et al.,38 the type of retention and

    bracket base are factors that influences the SBS. To the best of our knowledge, there

    are no studies involving SBS of Radiance® brackets.

  • 33

    The Inspire Ice® brackets presented mean SBS values of 16.09 ± 4.16 MPa,

    similar values to those of Theodorakopoulou et al.30 and close to those of Liu et al.36,

    who obtained mean values of 20.32 ± 8.00 MPa and 11.95 ± 4.60 MPa, respectively,

    using Inspire® brackets. However, it is worth emphasizing that Inspire Ice® brackets

    are a development of Inspire, with reduction in the retention area with the aim of

    diminishing the SBS. This is because bonding forces above 13.73 MPa should be

    avoided to eliminate the possibility of enamel fractures during bracket debonding.39,40

    To maintain the integrity of the dental enamel, the ideal adhesive failure

    should occur at the bracket-adhesive interface.38 The purpose of the ARI scores

    recommended by Årtun and Bergland13 is to quantify the adhesive remaining on the

    enamel, showing the location where the failure in the shear test occurred.

    In the metal bracket Miniature Twin®, the predominant ARI score was 1, with a

    frequency of 70%, with failure occurring at the adhesive-enamel interface; increasing

    the possibility of cracks in the enamel when it is debonded. In Clarity®, Radiance®

    and Inspire Ice® brackets adhesive failure occurred at the bracket-adhesive interface,

    with the great majority presenting an ARI score of 3; that is, all the adhesive

    remained adhered on the tooth with a frequency of 80%, 75% and, 65%,

    respectively. These results are in agreement with other studies that used ceramic

    brackets, which presented the advantage of maintaining enamel integrity and the

    disadvantage of requiring mechanical removal of the remaining adhesive.30-33

    The differences in the SBS results of brackets in this study, when compared

    with other studies,31-33 are probably due to the different methodologies and

    substrates used. This study showed that although the SBS of ceramic brackets was

    higher than that of metal brackets, adhesive failure occurred at the bracket-adhesive

  • 34

    interface, minimizing the chances of damage to the enamel. This makes the ceramic

    brackets a feasible and safe alternative for clinical use.

    CONCLUSIONS

    • Radiance® and Inspire Ice® brackets presented shear bond strength similar to

    metal brackets.

    • Clarity® brackets presented higher shear bond strength than the two

    monocrystalline ceramic brackets studied.

    • The monocrystalline ceramic brackets showed failure pattern at the bracket-

    adhesive interface, which suggests a tendency for the adhesive to remain

    adhered on the tooth.

  • 35

    REFERENCES

    1. Chaconas S, Caputo A, Niu G. Bond strength of ceramic brackets with various

    bonding systems. Angle Orthod 1990; 61:35-42.

    2. Britton JC, McInnes P, Weinberg R, Ledoux WR, Retief DH. Shear bond strength

    of ceramic orthodontic brackets to enamel. Am J Orthod Dentofacial Orthop 1990;

    98:348-53.

    3. Karamouzos A, Athanasios E, Papadopoulos MA. Clinical characteristics and

    properties of ceramic brackets: A comprehensive review. Am J Orthod Dentofacial

    Orthop 1990; 112:34-40.

    4. Blalock KA, Powers JM. Retention capacity of the bracket bases of new esthetic

    orthodontic brackets. Am J Orthod Dentofacial Orthop 1995; 107:596-603.

    5. Bordeaux JM, Moore RN, Bagby MD. Comparative evaluation of ceramic bracket

    base designs. Am J Orthod Dentofacial Orthop 1994; 105:552-60.

    6. Swartz ML. Ceramic brackets. J Clin Orthod 1988; 22:82-8.

    7. Omana HM, Moore RN, Bagby MD. Frictional properties of metal and ceramic

    brackets. J Clin Orthod 1992; 26:425-432.

    8. Birnie D. Ceramic brackets. Br J Orthod 1990; 17:71-75.

    9. Bishara SE, Fehr DE, Jacobsen JR. A comparative study of the debonding

    strengths of different ceramic brackets, enamel conditioners, and adhesives. Am

    J Orthod Dentofacial Orthop 1993;104:170-9.

    10. Wang WN, Meng CL, Tarng TH. Bond strength: a comparison between chemical

    coated and mechanical interlock bases of ceramic and metal brackets. Am J

    Orthod Dentofacial Orthop 1997; 111:374-81.

    11. Bishara SE. Ceramic brackets and the need to develop national standards. Am J

    Orthod Dentofacial Orthop 2000; 117:595-597

  • 36

    12. International Organization for Standardization. Technical Specification. Dental

    Materials – Testing of adhesion to tooth structure. ISO/TS 11405, Second

    edition, 2003 (E).

    13. Årtun J, Bergland S. Clinical trials with crystal growth conditioning as an

    alternative to acid-etch enamel pretreatment. Am J Orthod 1984; 85:333-40.

    14. Smith H, Casko J, Leinfelder K, Utley JD. Comparison of orthodontic bracket

    bond strengths: human vs. bovine enamel. J Dent Res 1976; 55:153-67.

    15. Nakamichi I, Iwaku M, Fusayama T. Bovine teeth as possible substitutes in the

    adhesion test. J Dent Res 1983; 62:1076-81.

    16. Oesterle L, Shellhart W, Belanger G. The use of bovine enamel in bonding

    studies. Am J Orthod Dentofacial Orthop 1998; 114:514-9.

    17. Iwamoto H, Kawamoto T, Kinosita Z. Bond strength of new ceramic brackets as

    studied in vitro. J Dent Res 1987; 66:928.

    18. Viazis A D, Cavanaugh G, Bevis RR. Bond strengths of ceramic bracket under

    shear stress: An in vitro report. Am J Orthod Dentofacial Orthop 1990;98:214-21.

    19. Redd TB, Shivapuja PK. Debonding ceramic brackets: effects on enamel. J Clin

    Orthod 1991; 25:475-81.

    20. Guess MB, Watanabe LG, Beck FM, Crall MG. The effect of silane coupling

    agents on the bond strength of a polycrystalline ceramic bracket. J Clin Orthod

    1988; 22:788-92.

    21. Gwinnett AJ. A comparison of shear bond strengths of metal and ceramic

    brackets. Am J Orthod Dentofacial Orthop 1988; 93:346-8.

    22. Merrill SW, Oesterle LJ, Hermesch CB. Ceramic bracket bonding: a comparison

    of shear, tensile, and torsional bond strengths of ceramic brackets. Am J Orthod

    Dentofacial Orthop 1994;106:290-7.

  • 37

    23. Ødegaard J, Segner D. Shear bond strength of metal brackets compared with a

    new ceramic bracket. Am J Orthod Dentofacial Orthop 1988; 94:201-6.

    24. Zachrisson BU. Bonding in Orthodontics. In: Graber TM, Swain BF, editors.

    Orthodontic, current principles and techniques. St Louis: Mosby; 2005.

    25. Chay SH, Wong SL, Mohamed N, Chia A, Yap AU. Effects of surface treatment

    and aging on the bond strength of orthodontic brackets to provisional materials.

    Am J Orthod Dentofacial Orthop 2007;132:577.e7-577.e11.

    26. Reynolds IR. A review of direct orthodontic bonding. Br J Orthod 1975; 2:171-8.

    27. Joseph VP, Rossouw E. The shear bond strengths of stainless steel and ceramic

    brackets used with chemically and lightactivated composite resins. Am J Orthod

    Dentofacial Orthop 1990; 97:121-5.

    28. Harris AMP, Joseph VP, Rossouw E. Comparison of shear bond strengths of

    orthodontic resins to ceramic and metal brackets. J Clin Orthod 1990; 24:725-8.

    29. Weinberger SJ, Foley TF, McConnell RJ, Wright GZ. Bond strengths of two

    ceramic brackets using argon laser, light, and chemically cured resin systems.

    Angle Orthod 1997; 3:173-8.

    30. Theodorakopoulou LP, Sadowsky PL, Jacobson A, Lacefield WJ. Evaluation of

    the debonding characteristics of 2 ceramic brackets: An in vitro study. Am J

    Orthod Dentofacial Orthop 2004; 125:329-36.

    31. Bishara SE, Olsen ME, Von Wald. Evaluation of debonding characteristics of a

    new collapsible ceramic bracket. Am J Orthod Dentofacial Orthop 1997; 112:

    552-9.

    32. Bishara SE, Olsen ME, Von Wald L, Jakobsen JR. Comparison of the debonding

    characteristics of two innovative ceramic bracket designs. Am J Orthod

    Dentofacial Orthop 1999; 116:86-92.

  • 38

    33. Mundstock KS, Sadowsky PL, Lacefield W, Bae S. An in vitro evaluation of a

    metal reinforced orthodontic ceramic bracket. Am J Orthod Dentofacial Orthop

    1999; 116:635-41.

    34. Yamamoto A, Yoshida T, Tsubota K, Takamizawa T, Kurokawa H, Miyazaki M.

    Orthodontic bracket bonding: Enamel Bond strength vs time. Am J Orthod

    Dentofacial Orthop 2006; 130:435.e1-435.e6.

    35. Oesterle LJ, Shellhart WC. Effect of aging on the shear bond strength of

    orthodontic brackets. Am J Orthod Dentofacial Orthop 2008;133:716-20.

    36. Liu JK, Chung CH, Chang CY, Shieh DB. Bond strength and debonding

    characteristics of a new ceramic bracket. Am J Orthod Dentofacial Orthop 2005;

    128:761-5.

    37. Ghafari J, Skanky T, Mante F. Shear bond strength of two ceramic brackets. J

    Clin Orthod 1992; 24: 491-3.

    38. Sinha PK, Nanda RS, Duncanson MG, Hosier MJ. Bond strengths and remnant

    adhesive resin on debonding for orthodontic bonding techniques. Am J Orthod

    Dentofacial Orthop 1995; 108:302-7.

    39. Retief DH. Failure at the dental adhesive etched enamel interface. J Oral Rehab

    1974; 1:265-84.

    40. Bishara SE, Fehr DE. Ceramic brackets: something old, something new, a

    review. Semin Orthod 1997; 3:178-88.

  • 39

    TABLES

    Table I. Descriptive statistics in Megapascals (MPa) of shear bond strengths of groups studied

    Groups n Mean Median Standard Deviation Games-Howell

    Miniature Twin® 20 13.245 11.110 5.155 a, c Clarity® 20 21.006 21.101 5.090 b Radiance® 20 12.081 11.984 2.139 a Inspire Ice® 20 16.096 16.184 4.164 c Equal letters means that there is no statistical difference among groups (P

  • 40

    FIGURES

    A1 B1

    Fig. 1. Images of brackets used in the study and their respective bases as seen by SEM (24 X). A1; A2, Miniature Twin®; B1; B2, Clarity®; C1; C2, Radiance® and D1; D2, Inspire Ice®.

    D2

    A2 B2 C2

    Fig. 2. Sequence of mounting test specimens. A, B, device with tooth fixed to 0.021” X 0.025” wire with elastic ligature; C, metallic inclusion ring placed for filling chemically cure acrylic resin; D, filling with acrylic; E, test specimen with bracket perpendicular to base and parallel to shear force direction and F, shear bond strength test.

    E F

    B A C

    D

    C1 D1

  • 41

    A D C B

    Fig. 3. SEM images (24x) of dental surface with representative ARI of each group. A, Miniature Twin® ARI 1; B, Clarity® - ARI 3; C, Radiance® - ARI 3 and D, Inspire Ice® - ARI 3.

  • 42

    3 ANEXOS

  • 43

    ANEXO 1- PARECER DO COMITÊ DE ÉTICA NO USO DE ANIMAIS

  • 44

  • 45

    ANEXO 2 – MATERIAL E MÉTODOS

    Fig 1. Seqüência da colagem dos brackets do grupo controle. A, Incisivo inferior bovino; B, profilaxia; C, condicionamento ácido; D, aplicação do adesivo; E, remoção dos excessos de resina com sonda exploradora, mantendo a pressão de 400gf com o dinamômetro; F, fotopolimerização; G, bracket colado com o slot na vertical, no longo eixo do dente.

    A B C D

    F G E

    Fig 2. Sistema adesivo Transbond™ XT (3M Unitek®, Monrovia, Ca). A, Transbond™ XT Primer; B, compósito Transbond™ XT.

    B A

  • 46

    Fig. 3. Brackets utilizados. A, Miniature Twin®; B, Clarity®; C, Radiance®, D, Inspire Ice®.

    B C D A

    Fig 4. Seqüência da montagem dos corpos de prova. A; B, dispositivo para a inclusão dos dentes; C; D, dente preso no fio 0,021” x 0,025” com ligadura elástica; E; F; G, anel metálico de inclusão posicionado para o preenchimento com acrílico autopolimerizável; H, preenchimento com o acrílico; I; J, corpo de prova com o bracket perpendicular à base e paralelo à direção da força de cisalhamento; K, máquina universal de ensaios EMIC® DL 500; e L, ensaio da RA ao cisalhamento.

    I J K L

    A B C D

    G H E F

  • 47

    ANEXO 3 – ANÁLISE ESTATÍSTICA

    TABELA I. TESTE DE NORMALIDADE KOLMOGOROV- SMIRNOV PARA OS GRUPOS ESTUDADOS, PUCPR - 2008

    Grupos Estatística n Valor p

    Miniature Twin® 0,196 20 0,043*

    Clarity® 0,109 20 0,200

    Radiance® 0,102 20 0,200

    Inspire Ice® 0,132 20 0,200 FONTE: dados da pesquisa NOTA: distribuição normal para p>0,05.

    TABELA II. TESTE DE HOMOGENEIDADE DE VARIÂNCIA DE LEVENE

    PARA VARIÁVEL RESISTÊNCIA ADESIVA, PUCPR - 2008

    Variável Estatística g.l.1 g.l.2 Valor p

    Resistência Adesiva 3,436 3 76 0,021* FONTE: dados da pesquisa LEGENDA: g.l. graus de liberdade NOTA: homogeneidade de variância para p>0,05. TABELA III. VALORES DE p DE ACORDO COM O TETE DE

    COMPARAÇÕES MÚLTIPLAS GAMES-HOWELL, PUCPR - 2008

    Miniature Twin® Clarity® Radiance® Inspire Ice®

    Miniature Twin® 0,000* 0,788 0,236

    Clarity® 0,000* 0,000* 0,010*

    Radiance® 0,788 0,000* 0,003*

    Inspire Ice® 0,236 0,010* 0,003* FONTE: dados da pesquisa NOTA: * indica diferença estatística (p

  • 48

    Fig 1. Resistência adesiva média (MPa), segundo grupos.

    Fig 2. Prevalência do índice de adesivo Remanescente (IAR), segundo grupos.

  • 49

    ANEXO 4 - NORMAS DO AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS Information for Authors Electronic manuscript submission and review The American Journal of Orthodontics and Dentofacial Orthopedics now uses Editorial Manager, an online manuscript submission and review system. To submit or review an article, please go to the AJO-DO Editorial Manager website: ees.elsevier.com/ajodo . Send other correspondence to: Dr David L. Turpin, DDS, MSD, Editor-in-Chief American Journal of Orthodontics and Dentofacial Orthopedics University of Washington Department of Orthodontics, D-569 HSC Box 357446 Seattle, WA 98195-7446 Telephone (206)221-5413 Fax (206)221-5467 E-mail: [email protected] General Information The American Journal of Orthodontics and Dentofacial Orthopedics publishes original research, reviews, case reports, clinical material, short communications, and other material related to orthodontics and dentofacial orthopedics. Submitted manuscripts must be original, written in English, and not published or under consideration elsewhere. Manuscripts will be reviewed by the editor and consultants and are subject to editorial revision. Authors should follow the guidelines below. Statements and opinions expressed in the articles and communications herein are those of the author(s) and not necessarily those of the editor(s) or publisher, and the editor(s) and publisher disclaim any responsibility or liability for such material. Neither the editor(s) nor the publisher guarantees, warrants, or endorses any product or service advertised in this publication; neither do they guarantee any claim made by the manufacturer of any product or service. Each reader must determine whether to act on the information in this publication, and neither the Journal nor its sponsoring organizations shall be liable for any injury due to the publication of erroneous information. Guidelines for Original Articles Submit Original Articles via the online Editorial Manager: ees.elsevier.com/ajodo . Organize your submission as follows. 1. Title Page. Put all information pertaining to the authors in a separate document. Include the title of the article, full name(s) of the author(s), academic degrees, and institutional affiliations and positions; identify the corresponding author and include an address, telephone and fax numbers, and an e-mail address. This information will not be available to the reviewers. 2. Abstract. Structured abstracts of 200 words or less are preferred. A structured abstract contains the following sections: Introduction, describing the problem; Methods, describing how the study was performed; Results, describing the primary results; and Conclusions, reporting what the authors conclude from the findings and any clinical implications. 3. Manuscript. The manuscript proper should be organized in the following sections: Introduction and literature review, Material and Methods, Results, Discussion, Conclusions, References, and figure captions. Express measurements in metric units whenever practical. Refer to teeth by their full name or their FDI tooth number. For style questions, refer to the AMA Manual of Style, 9th edition. Cite references selectively, and number them in the order cited. Make sure that all references have been mentioned in the text. Follow the format for references in "Uniform Requirements for Manuscripts Submitted to Biomedical Journals" (Ann Intern Med 1997;126:36-47); http://www.icmje.org . Include the list of references with the manuscript proper. Submit figures and tables separately (see below); do not embed figures in the word processing document.

  • 50

    4. Figures. Digital images should be in TIF or EPS format, CMYK or grayscale, at least 5 inches wide and at least 300 pixels per inch (118 pixels per cm). Do not embed images in a word processing program. If published, images could be reduced to 1 column width (about 3 inches), so authors should ensure that figures will remain legible at that scale. For best results, avoid screening, shading, and colored backgrounds; use the simplest patterns available to indicate differences in charts. If a figure has been previously published, the legend (included in the manuscript proper) must give full credit to the original source, and written permisson from the original publisher must be included. Be sure you have mentioned each figure, in order, in the text. 5. Tables. Tables should be self-explanatory and should supplement, not duplicate, the text. Number them with Roman numerals, in the order they are mentioned in the text. Provide a brief title for each. If a table has been previously published, include a footnote in the table giving full credit to the original source. 6. Model release and permission forms. Photographs of identifiable persons must be accompanied by a release signed by the person or both living parents or the guardian of minors. Illustrations or tables that have appeared in copyrighted material must be accompanied by written permission for their use from the copyright owner and original author, and the legend must properly credit the source. Permission also must be obtained to use modified tables or figures. 7. Copyright release. In accordance with the Copyright Act of 1976, which became effective February 1, 1978, all manuscripts must be accompanied by the following written statement, signed by all authors: "The undersigned author(s) transfers all copyright ownership of the manuscript [insert title of article here] to the American Association of Orthodontists in the event the work is published. The undersigned author(s) warrants that the article is original, does not infringe upon any copyright or other proprietary right of any third party, is not under consideration by another journal, has not been previously published, and includes any product that may derive from the published journal, whether print or electronic media. I (we) sign for and accept responsibility for releasing this material." Scan the printed copyright release and submit it via the Editorial Manager, or submit it via fax or mail. 8. Conflict of interest statement. Report any commercial association that might pose a conflict of interest, such as ownership, stock holdings, equity interests and consultant activities, or patent-licensing situations. If the manuscript is accepted, the disclosed information will be published with the article. The usual and customary listing of sources of support and institutional affiliations on the title page is proper and does not imply a conflict of interest. Guest editorials, Letters, and Review articles may be rejected if a conflict of interest exists. Other Articles Follow the guidelines above, with the following exceptions, and submit via Editorial Manager. Case Reports will be evaluated for completeness and quality of records, quality of treatment, uniqueness of the case, and quality of the manuscript. A high quality manuscript will include the following sections: introduction; diagnosis; etiology; treatment objectives, alternatives, progress, and results; and discussion. The submitted figures should include extra oral and intraoral photographs and dental models, panoramic radiographs and tracings from both pretreatment and posttreatment, and progress or retention figures as appropriate. Short Communications should not exceed 2000 words, including the bibliography, and should include a minimal number of figures or tables. Priority will be given to communications relating to primary research data, preferably clinical but also basic. This section permits time-sensitive material to be published within 6 months of submission. Techno Bytes items report on emerging technological developments and products for use by orthodontists. Litigation, Legislation, and Ethics items report legal and ethical issues of interest to orthodontists. Miscellaneous Submissions Letters to the Editor and Ask Us questions and answers appear in the Readers' Forum section and are encouraged to stimulate healthy discourse concerning the profession. Send letters or questions directly to the editor, via e-mail: [email protected]. Submit a signed copyright release with the letter, or fax or mail separately. Brief, substantiated commentary on subjects of interest to the orthodontic profession is occasionally published as a Guest Editorial or Special Article. Send Guest Editorials or Special Articles directly to

  • 51

    the editor, via e-mail: [email protected]. Submit a signed copyright release with the editorial, or fax or mail separately. Books and monographs (domestic and foreign) will be reviewed, depending on their interest and value to subscribers. Send books to the Editor of Reviews and Abstracts, Dr Alex Jacobson, University of Alabama School of Dentistry, 1919 7th Ave S, Box 23, Birmingham, AL 35294. They will not be returned. Checklist for authors ____Title page, including full name, academic degrees, and institutional affiliation and position of each author, and author to whom correspondence and reprint requests are to be sent, including address, business and home phone numbers, fax numbers, and e-mail address ____Abstract ____Article proper, including references and figure legends ____Figures, in TIF or EPS format ____Tables ____Copyright release statement, signed by all authors ____Photographic consent statement(s) ____Conflict of interest statement ____Permissions to reproduce previously published material

  • Livros Grátis( http://www.livrosgratis.com.br )

    Milhares de Livros para Download: Baixar livros de AdministraçãoBaixar livros de AgronomiaBaixar livros de ArquiteturaBaixar livros de ArtesBaixar livros de AstronomiaBaixar livros de Biologia GeralBaixar livros de Ciência da ComputaçãoBaixar livros de Ciência da InformaçãoBaixar livros de Ciência PolíticaBaixar livros de Ciências da SaúdeBaixar livros de ComunicaçãoBaixar livros do Conselho Nacional de Educação - CNEBaixar livros de Defesa civilBaixar livros de DireitoBaixar livros de Direitos humanosBaixar livros de EconomiaBaixar livros de Economia DomésticaBaixar livros de EducaçãoBaixar livros de Educação - TrânsitoBaixar livros de Educação FísicaBaixar livros de Engenharia AeroespacialBaixar livros de FarmáciaBaixar livros de FilosofiaBaixar livros de FísicaBaixar livros de GeociênciasBaixar livros de GeografiaBaixar livros de HistóriaBaixar livros de Línguas

    http://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.brhttp://www.livrosgratis.com.br/cat_1/administracao/1http://www.livrosgratis.com.br/cat_1/administracao/1http://www.livrosgratis.com.br/cat_1/administracao/1http://www.livrosgratis.com.br/cat_1/administracao/1http://www.livrosgratis.com.br/cat_1/administracao/1http://www.livrosgratis.com.br/cat_1/administracao/1http://www.livrosgratis.com.br/cat_1/administracao/1http://www.livrosgratis.com.br/cat_2/agronomia/1http://www.livrosgratis.com.br/cat_2/agronomia/1http://www.livrosgratis.com.br/cat_2/agronomia/1http://www.livrosgratis.com.br/cat_2/agronomia/1http://www.livrosgratis.com.br/cat_2/agronomia/1http://www.livrosgratis.com.br/cat_2/agronomia/1http://www.livrosgratis.com.br/cat_2/agronomia/1http://www.livrosgratis.com.br/cat_3/arquitetura/1http://www.livrosgratis.com.br/cat_3/arquitetura/1http://www.livrosgratis.com.br/cat_3/arquitetura/1http://www.livrosgratis.com.br/cat_3/arquitetura/1http://www.livrosgratis.com.br/cat_3/arquitetura/1http://www.livrosgratis.com.br/cat_3/arquitetura/1http://www.livrosgratis.com.br/cat_3/arquitetura/1http://www.livrosgratis.com.br/cat_4/artes/1http://www.livrosgratis.com.br/cat_4/artes/1http://www.livrosgratis.com.br/cat_4/artes/1http://www.livrosgratis.com.br/cat_4/artes/1http://www.livrosgratis.com.br/cat_4/artes/1http://www.livrosgratis.com.br/cat_4/artes/1http://www.livrosgratis.com.br/cat_4/artes/1http://www.livrosgratis.com.br/cat_5/astronomia/1http://www.livrosgratis.com.br/cat_5/astronomia/1http://www.livrosgratis.com.br/cat_5/astronomia/1http://www.livrosgratis.com.br/cat_5/astronomia/1http://www.livrosgratis.com.br/cat_5/astronomia/1http://www.livrosgratis.com.br/cat_5/astronomia/1http://www.livrosgratis.com.br/cat_5/astronomia/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_6/biologia_geral/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_8/ciencia_da_computacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_9/ciencia_da_informacao/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_7/ciencia_politica/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_10/ciencias_da_saude/1http://www.livrosgratis.com.br/cat_11/comunicacao/1http://www.livrosgratis.com.br/cat_11/comunicacao/1http://www.livrosgratis.com.br/cat_11/comunicacao/1http://www.livrosgratis.com.br/cat_11/comunicacao/1http://www.livrosgratis.com.br/cat_11/comunicacao/1http://www.livrosgratis.com.br/cat_11/comunicacao/1http://www.livrosgratis.com.br/cat_11/comunicacao/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_12/conselho_nacional_de_educacao_-_cne/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_13/defesa_civil/1http://www.livrosgratis.com.br/cat_14/direito/1http://www.livrosgratis.com.br/cat_14/direito/1http://www.livrosgratis.com.br/cat_14/direito/1http://www.livrosgratis.com.br/cat_14/direito/1http://www.livrosgratis.com.br/cat_14/direito/1http://www.livrosgratis.com.br/cat_14/direito/1http://www.livrosgratis.com.br/cat_14/direito/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_15/direitos_humanos/1http://www.livrosgratis.com.br/cat_16/economia/1http://www.livrosgratis.com.br/cat_16/economia/1http://www.livrosgratis.com.br/cat_16/economia/1http://www.livrosgratis.com.br/cat_16/economia/1http://www.livrosgratis.com.br/cat_16/economia/1http://www.livrosgratis.com.br/cat_16/economia/1http://www.livrosgratis.com.br/cat_16/economia/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_17/economia_domestica/1http://www.livrosgratis.com.br/cat_18/educacao/1http://www.livrosgratis.com.br/cat_18/educacao/1http://www.livrosgratis.com.br/cat_18/educacao/1http://www.livrosgratis.com.br/cat_18/educacao/1http://www.livrosgratis.com.br/cat_18/educacao/1http://www.livrosgratis.com.br/cat_18/educacao/1http://www.livrosgratis.com.br/cat_18/educacao/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_19/educacao_-_transito/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_20/educacao_fisica/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_21/engenharia_aeroespacial/1http://www.livrosgratis.com.br/cat_22/