Cromatografia Líquida Acoplada àEspectrometria de Massas (Sequencial)
Técnicas de Ionização eAnalisadores de Massas
Universidade Federal de Santa Maria (UFSM) Departamento de Química Campus da UFSM 97105-900 Santa Maria - RSProf. Dr. Renato Zanella e-mail: [email protected]
2011
Mass Spectrometry
2
LiquidChromatography Ionization Mass Analyzer
Detector/Data
Collection
•ESI•APCI•APPI
•Triple Quadrapoles•Ion-Traps•Hybrids
•Very important!•Many columns•Many solvent systems
3
Em LCMS/MS (MRM) a intensidade do pico do analito é baixa se comparada a uma análise LCMS realizada no modo SIM (Single Ion Monitoring). No entanto, o LCMS/MS fornece maior sensibilidade que o LCMS devido à melhora significativa dos níveis de ruído com consequente aumento da relação sinal/ruído.
4
Mass Analyzers• Double Focusing Magnetic Sector• Quadrupole Mass Filter• Quadrupole Ion Trap• Linear Time-of-Flight (TOF)• Reflectron TOF• Fourier Transform Ion Cyclotron Resonance
(FT-ICR-MS)
5
Main Components of the LC-MS
Ion Optics Detector
Spray Chamber
⊕⊕⊕ ⊕⊕⊕
Mass Filter
6
Triple Quadrupole ConfigurationQ1 Q2 Q3Q0
RF only Collision Cell
Scanning RF/DC
ScanningRF/DC
RF only
• Q1 and Q3 are standard mass filter quadrupoles.– The can scan masses sequentially (e.g. 50 to 500 amu)– The can be used to select a single mass.
• Q2 is an RF only quadrupole that is in a gas filled chamber. – Q2 is the “collision cell” where mass fragmentation occurs. – Q2 does not filter ions. It accepts all ion sent to it by Q1 and
passes all ions formed by collision to Q3 to be sorted.
QqQ
7
Ao realizar medidas dos íons (MS/MS automático ou íons específicos) no modo MRM, espectros de massa do íon precursor podem ser obtidos durante uma transição MRM. A estrutura do composto pode ser simultaneamente confirmada pelo espectro do produto resultante do íon devido à alta seletividade durante o MRM e sem a interferência de coeluição com outras substâncias.
Cromatograma de padrões de 226 pesticidas obtidos em 2 minutos.
8
500 transições MRM em um segundo
Medida simultânea de íons produzidos no modo MRM.
9
MRM = monitoramento de múltiplas reações(Multiple Reaction Monitoring)
MRM é também chamado de SRM (Selected Reaction Monitoring)
Q1 Q2 Q3Q0
RF only Collision
Cell
Scanning RF/DC
ScanningRF/DC
RF only
10
API-Electrospray
GC/MS
ParticleBeam
APCI1000
100,000
10,000
Nonpolar Very polar
FABThermospray
Mol
ecul
ar W
eigh
t
Mass Spectrometry
11
Mass Spectrometry
Terms and Definitions
Molecular ion / precursor ionIon formed by ionization of the analyte species
Fragment ions / product ionsIons formed by the gas-phase dissociation of the molecular ion
Relative AbundanceRelative Abundance is a measure of the relative amount of ion signal recorded by the detector
12
13
Ionization Techniques
Gas-Phase Methods• Electron Impact (EI)• Chemical Ionization (CI)
Desorption Methods• Matrix-Assisted Laser Desorption Ionization (MALDI)• Fast Atom Bombardment (FAB)
Spray Methods• Electrospray (ESI)• Atmospheric Pressure Chemical Ionization (APCI)• Atmospheric Pressure Photo-Ionization (APPI)• New dual sources (ESI/APCI) or (APCI/APPI)
GC
14
Electron Impact GC
15
Electron Impact
Advantages• Well-Established• Fragmentation Libraries• No Supression• Insoluble Samples• Interface to GC• Non-Polar Samples
Disadvantages• Parent Identification• Need Volatile Sample• Need Thermal Stability• No Interface to LC• Low Mass Compounds
(<1000 amu)• Solids Probe Requires
Skilled Operator
(low picomole)
16
Chemical IonizationGC
17
Chemical Ionization
Advantages• Parent Ion• Interface to GC• Insoluble Samples
Disadvantages• No Fragment Library• Need Volatile Sample• Need Thermal Stability• Quantitation Difficult• Low Mass Compounds
(<1000 amu)• Solids Probe Requires
Skilled Operator
(low picomole)
18
Electrospray Ionization (ESI)
(Eletronebulização)
(up to 6 kV)
19
Electrospray
Ionization (ESI)
Typical ions produced by electrospray ionisation:
Positive mode:[M+H]+ protonated molecule[M+Na] +, [M+K] + … adducts[M+CH3CN+H] + protonated, + solvent adducts
Negative mode:[M-H] - deprotonated molecule[M+HCOO -] -, … adducts 20
Electrospray: Types of Ions Formed
• ESI can operate in either + or - mode.
Positive mode:– Best suited to basic drugs that form a stable HCl salt.
• [M+H]+ is the primary ion formed• [M+nH]n+ and [M+Na+]+ can also be formed.
Negative mode:– Best suited to acidic drugs that form stable Na salts.
• [M-H]-, [M-nH]n- and [M+I-]-
21
++
Heated nitrogen dryinggas
Dielectric capillary entrance
Nebulizer (gas shown in red)
Solvent spray
Electrospray Ions
-4000 V
⊕⊕
⊕⊕ ⊕
⊕⊕ ⊕⊕ ⊕⊕ ⊕ ⊕⊕ ⊕ ⊕
Evaporation RayleighLimit
CoulombExplosion
Evaporation Analyte Ion
API-Electrospray IonizationAtmospheric Pressure Ionization
22
Increasing fragmentation with increasing cone voltage. The fragmentation obtained by in source fragmentation is useful for confirming a peak identity, getting structure information, improving selectivity, building libraries…
23
ESI = Ionização por eletronebulização à pressão atmosférica
1º) Bombeamento da amostra através de uma agulha nebulizadora;
2º) Aplicação de uma voltagem entre a agulha e o cone que a rodeia;
3º) Formação de uma nuvem de gotículas carregadas;
4º) Evaporação do solvente devido à presença de um dispositivo aquecido;
5º) Aumento da densidade de cargas na superfície das gotículas;
6º) Repulsão entre as cargas até os íons evaporarem das gotículas.
Electrospray Ionization (ESI)
24
ESI
Advantages• Parent Ion• High Mass Compounds
(>100,000 amu)• Thermally Labile
Compounds (<0º C)• Easy to Operate• Interface to HPLC• Zeptomole sensitivity
with nanospray
Disadvantages• No Fragmentation• Need Polar Sample• Need Solubility in Polar
Solvent (MeOH, ACN, H2O, Acetone are best)
• Sensitive to Salts• Supression
(low femtomole to zeptomole)
25
Atmospheric PressureChemical Ionization (APCI)
Solvent molecules (S) being protonated by the corona (SH+), thenreacting with the analyte molecule (M) to give the protonated form MH+.
Atmospheric Pressure Ionization
26
Atmospheric Pressure Chemical Ionization (APCI)
APCI = Ionização Química à Pressão Atmosférica
1º) Bombeamento da amostra através de um capilar;
2º) Vaporização do solvente e da amostra, por um dispositivo aquecido;
3º) Ocorre “descarga elétrica em coroa”;
4º) Ionização do solvente da fase móvel;
5º) Os íons formados reagem com as moléculas do analito.
27
APCI
Advantages• Parent Ion• Insensitive to Salts• Interface to HPLC• Can use Normal Phase
Solvents• Handles High Flow Rates
Disadvantages• Need Volatile Sample• Need Thermal Stability
(high femtomole)
28
Seleção da fonte:• Experiência do laboratório e o tipo de instrumento;
• ESI é mais frequentemente utilizada;• Facilidade do uso;• Robustez;• Seletividade para mais compostos
• APCI: • Máxima sensibilidade no modo íon positivo para compostos neutros;
• Para compostos de íon negativo que requer captura de elétron para ionização (nitratos)
APCI : N-metilcarbamatos, compostos neutros
ESI : • técnica preferida e menos propensa a degradação térmica• trabalha numa faixa mais ampla de compostos
29
Principais Diferenças entre APCI e ESI• Mecanismo de Ionização:– Enquanto ESI tem voltagem aplicada à ponta do spray, APCI apresenta um nebulizador aquecido e pneumaticamente assistido. A voltagem (~3 kV) é aplicada a uma agulha de metal na saída do spray.– A descarga Corona ioniza as moléculas do solvente que por sua vez ionizam os analitos por transferência de prótons formando [M+H]+ ou [MH]-
• Vazão da FM:– APCI tolera vazões na ordem de 0,2 a 2,0 mL/min, enquanto que ESI opera no máximo até1,0 mL/min.– ESI é melhor indicado para fluxos tão baixos quanto 5 μL/min, compatível com acoplamentos capilares (μLC ou CZE).
• Fragmentação:– Devido ao uso de aquecimento, APCI pode produzir alguma fragmentação, enquanto que ESI pode até formar alguns íons pseudo-moleculares.– APCI não produz cargas múltiplas, portanto não é adequado para compostos de alta MM.
• Sensibilidade:– Sem ser uma regra, APCI tende a render melhor sensibilidade para solutos menos polares.
30
ESI characteristicsSoft ionization method, provides molecular weight information.
Suitable for analyzing large bio- or synthetic polymers.
Suitable for analyzing polar and even ionic compounds.
Less fragmentation. APCI characteristics
Provides molecular weight information.
Suitable for analyzing less polar compounds compared to ESI.
Increased fragmentation compared to ESI.
MALDI characteristicsSoft ionization method, provides molecular weight information.
Suitable for analyzing very large bio- or synthetic polymers.
Suitable for analyzing polar and even ionic compounds.
Less fragmentation.
Ionisation Techniques
31
ESI e APCI
32
Priority pollutant PAHs in surface and drinking water by LC/APPI/MS-MS
Two specific transitions per compound were selected for simultaneous monitoring in SRM mode.
LC-MS/MS SRM chromatograms of a standard solution (100 μg L-1; 1000 μg L-1 for Naphth, Fluor)
The photoionization mechanism is simplified under vacuum conditions: photon absorption by the analyte molecule, leading to electron ejection, forming a molecular radical cation M.+
Atmospheric pressure photoionization (APPI)
Photoionization
Benefits of Photoionization� Ionizes wide range of compounds (e.g.,
non-polars, electronegative cpds, etc.)� Predominantly parent ion signal� Minimum fragmentation� Minimum solvent signal� Minimum ion suppression� Signal linear with concentration
H2O, CO2, O2, N2
Ionization potential
Drugs
Chemicalweapons
Aromatics
MeOH, AcCN, chloro-solvents
Solvent (S) Analyte (A)
Ener
gy [e
V]
IP
IP
AA++
SS
[A[A--m]m]++ + m+ m
Fragmentation
33
APPI Source
34
Direct APPI vs. Dopant-assisted APPIDirect APPI
M + hv → M+ + e-
M+ + S → MH+ + S[-H]
Dopant APPI
D + hv → D+ + e-
D+ + M → MH+ + D[-H]D+ + M → M+ + D
Analyte molecule M is ionized to a molecular radical ion M+. (If analyte ionization potential is below photon energy)
In the presence of protic solvents, M+ may abstract a hydrogen atom to form MH+.
A photoionizable dopant is delivered in large concentration to yield many D+ ions.
D+ ionizes analyte M by proton or electron transfer.
This is PI-initiated APCI35 36
Which is Best ionization source?
• It depends on the exact application.• Increasing polarity and molecular weight and thermal
instability favors ESI.• Most drugs of abuse are highly polar and are easily
analyzed using ESI.• High molecular weight proteins also require ESI• Lower polarity and molecular weight favors APCI or APPI.• Lower background, but compounds must be more
thermally stable.
37
Fast Atom Bombardment (FAB)
38
FAB
Advantages• Parent Ion• High Mass Compounds
(10,000 amu)• Thermally Labile
Compounds (R.T.)
Disadvantages• No Fragment Library• Solubility in Matrix (MNBA,
Glycerol)• Quantitation Difficult• Needs Highly Skilled
Operator• Relatively Low Sensitivity
(nanomole)
39
Matrix-Assisted Laser Desorption Ionization (MALDI)
http://www.noble.org/PlantBio/MS/ion_tech_main.html 40
MALDI
Advantages• Parent Ion• High Mass Compounds
(>100,000 amu)• Thermally Labile
Compounds (R.T.)• Easy to Operate
Disadvantages• No Fragment Library• Wide variety of matrices• Quantitation Difficult
(low femtomole)
41
Data Acquisition ModesSIM vs SRM
Ø Selected-Ion MonitoringIf the analysis aims at detecting target compounds of known
spectral characteristics with maximum sensitivity
Ø Selected-Reaction MonitoringDetection of selected reactions, based on the decomposition
reactions of ions that are characteristic of the compounds to be analysed
Three Acquisition Modes:
Ø ScanningComplete spectra are repeatedly measured between two
extreme masses
42
Double-Focusing Magnetic Sector
http://www.asms.org
Advantages• Very High Resolution (60,000)• High Accuracy (<5 ppm)• 10,000 Mass Range
Disadvantages• Very Expensive• Requires Skilled Operator• Difficult to Interface to ESI• Low resolution MS/MS without
multiple analyzers
Analisadores
43
Analisador Quadrupolar (Q)• Analisador de varredura• Conjunto de 4 pólos de sinais opostos, que alternam sinais de radiofreqüência entrepares, permitindo que apenas um íon atinja o detector de cada vez
Analisadores
44
*
*Negative Rods
Positive Rods
Positive Rods
Filtering High Mass
Filtering Selected Mass
Filtering Low Mass
M+
M+
M+
How Does a Quadrupole Mass Filter Work?
Filters out all m/z values except the one it is set to passObtains a mass spectrum by sweeping across the entire mass range
Analisadores
45
MS/MS with Triple QuadrupolesThe analyser of a “triple quad” instrument consists in two quadrupoles, separated by a collision cell. Such a configuration is often referred as a "tandem in space" instrument. Precursor ions and product ions are created and analysed in different physical spaces.
A triple quad instrument can be used in various ways, which are represented on the next figure. The firsts two experiments are in fact using the triple quad in a single quadrupole mode
Analisadores
46
MS/MS with Triple Quadrupoles
- The first quad (Q1) is set to pass one ion of interest- The second quad (q2) is set to pass all ions and is flooded with a collision gas to fragment the
parent ion present (collision induced dissociation or CID)- The third quad (Q3) obtains a daughter ion spectrum by sweeping across the desired mass range
Analisadores
47
• Os íons são acelerados para aumentar a energia cinética e passar em uma câmara com o gás de colisão.
Usa-se, geralmente: Ar para o Triplo Quadrupolo
He para a Ion Trap
• o uso da fonte CID tem desvantagens importantes, como a dificuldade prática de funcionamento na voltagem ótima de fragmentação para todos compostos em um método multirresíduo.
• As desvantagens podem ser resolvidas trabalhando com MS/MS Sequencial pelo uso do Triplo Quadrupolo ou sistemas Íon Trap porque a voltagem usada pode ser mais baixa ou semelhante para todos os compostos e o número de fragmentos produzidos pode ser controlado por um bom ajuste do MS/MS para especificar a fragmentação e produzir espectros simples.
CID = Dissociação induzida por colisão Analisadores
48
Analisadores
49
Analisadores
50
Advantages• Inexpensive• Easily Interfaced to
Many Ionization Methods
Disadvantages• Low Resolution (<4000)• Low Accuracy (>100 ppm)• MS/MS requires multiple
analyzers• Low Mass Range (<4000)• Slow Scanning
Quadrupole Mass Analyzer/FilterAnalisadores
51
Ion-Trap Analyzer
• Acúmulo de íons no interior de um campo quadrupolar 3D;
• Liberação controlada de íons de diferentes massas;
• Permite realizar MSn
Analisadores
52
Ion-Trap Analyzer
MS
MS/MS
53
Advantages• Inexpensive• Easily Interfaced to
Many Ionization Methods
• MS/MS in one analyzer
Disadvantages• Low Resolution (<4000)• Low Accuracy
(>100ppm)• Space Charging
Causes Mass Shifts• Low Mass Range
(<4000)• Slow Scanning
Quadrupole Ion TrapAnalisadores
54
Espectrômetros de Massas em Sequência (MS/MS)
• Podem ser de dois modos:
• Consecutivos no espaço usando dois espectrometros separados;
• Sucessivos no tempo usando a mesma massa no sistema mais de uma vez (ion trap)
• A técnica permite a medida de fragmentação de um pico selecionado em um espectro de massa que produz o espectro de massas do íon precursor;
• A informação é aumentada com a combinação de dois MS;
Analisadores
55
Linear Time-of-Flight (TOF)
Advantages• Extremely High Mass Range (>1 MDa)• Fast Scanning
Disadvantages• Low Resolution (4000)• Low Accuracy (>200 ppm)• MS/MS not possible
• íons são produzidos em pulsos e acelerados em alta velocidade por um campo elétrico num tubo de deslocamento.• a distância da origem até o detector é fixa,
• o tempo que o íon leva para atravessar o analisador é inversamente proporcional a sua velocidade
Analisadores
56
Reflectron Time-of-Flight (TOF)
Advantages• High Resolution (>20,000 in some models)• High Accuracy (<5ppm)• 10,000 Mass Range• Fast Scanning
Disadvantages• Low Resolution for MS/MS
Analisadores
57
Z-sprayíon souce
Hexapoleíon bridge
Quadrupole Hexapolecollision cell
MS-MS
Reflectron
QUADRUPOLE MS TOF MS
Sistema Q-TOF/MSCombinação de um MS Quadrupolo com um analisador TOF
• Espectros de massas contínuos “full-scan” tornam o MS/MS Q TOF mais sensível (10 - 100 vezes) que os Quadrupolos em sequência;
• TOF e o MS/MS Q TOF tem a mesma sensibilidade no modo scan e no modo de seleção do íon;
• A maior capacidade do MS/MS Q TOF comparado com instrumentos Triplo Quadrupolo e MS/MS Íon Trap está na sua habilidade para determinar massas precisas nos fragmentos de íons gerados na cela de colisão.
Algumas características do Q-TOF/MS
Analisadores
58
O ESPECTRO DE MASSAS
• Espectro de massas = diagrama da intensidade relativa dos picos em função das razões massa-carga.
9Pico base = pico com a intensidade mais alta, arbitrariamente assinalada em 100. Em LC-MS o pico base geralmente é a molécula protonada e corresponde ao íon positivo mais estável;
5Íon molecular = é o pico de maior razão massa-carga e representa o peso molecular do composto.
59
The API Mass SpectrumElectrospray (ESI) and APCI are both API (atmospheric pressure ionisation) techniques and produce a soft ionisation. Thus, the MS spectra obtained with API ionisation will consist mainly of the "molecular" ions, unless fragmentation techniques are applied. The possible fragmentation techniques are in source CID (collision induced dissociation), CID in the collision cell of a tandem type instrument, fragmentation in an ion trap. This is very different from the spectra obtained with EI (electron impact ionisation).
sulfamethazine spectrum Negative ESI mass spectrum for glucose, with chloride adducts. 60
Operating conditionsEluent: the mobile phase must be suitable for ionisation. If it is not the case, a small amount of modifier can be added, in order to generate the solvent gas phase ions.
Buffers: buffers must be volatile.
• Tampões voláteis (ex. Formiato de amônio)concentração do tampão, ácido ou base usado para ajustar o pH deveriam ser o mais baixa possível (0,1%);
Substitute phosphates, sulfates, and borates with ammonium acetate or formate, trifluoroacetic acid (TFA), heptafluorobutric acid (HFBA), tetrabutylammonium hydroxide (TBAH)
61
Choosing Between Ionization TechniquesGeneral rules can be given to chose between ESI, APCI, or EI.ESI is preferred for compounds which are ionic or very polar or thermo labile, or with masses higher than 1000…APCI is preferred for compounds which are not very polar.If we take the example of pesticides, we can rule out that: polar pesticides will be in ESI, less polar in APCI, volatiles in APCI, non polar better done with GC/EI
62
Mass DefinitionThe mass spectrometer measures the exact mass. Looking at the below mass spectra, the most abundant peak is at 221.95 (top) and 219.81 (bottom). These spectra are obtained with positive ionisation (top) and negative ionisation (bottom). The peaks correspond to the protonated or deprotonated molecule.
If we look for the molecular mass of the chloridazon pesticide, we can find various values in the spectra1.221.6379: this is the average mass. It is based on the average atomic masses. 2.221: this is the nominal mass, calculated on the nominal mass of the most abundant isotopes 3.221.0278: this is the exact (or monoisotopic) mass, based on the exact mass of the most abundant natural isotopesThe value is slightly different from the expected 222.0278 and 219.0278 because these spectra were obtained with a quadrupole instrument, which does not provide sufficient mass resolution and mass accuracy for obtaining the exact mass.The next smaller peaks correspond to the C13 and Cl37 isotopes.
63
Benzoylurea pesticides ESI -
64
Organophosphorus ESI +
65
Carbamate analysis by LC/MSColumn: Ultra Carbamate, 100 x 4.6 mm, 3 µm Conc.: 100 µg/L eachMobile Phase: A: 90:10 water:methanol + 10 mM ammonium formate
B: 10:90 acetonitrile:methanol + 10 mM ammonium formateGradient: 90:10 A:B to 10:90 A:B from 0-15 min.Flow-rate: 0.25mL/minMode: ESI+
66mass/charge
abundance
Base Peak: most abundant ion in spectrum
[M + H]+
Psuedomolecular ionor Quasi-molecular ion
A + 1A + 2
from naturally occurring isotopes{
Mass Spectra
67
Comparação entre instrumentos MS/MSInstrumento Sensibilidade
em full scanSeletividade Exatidão Faixa
dinamicaCaracterísticas
Triplo quadrupolo baixa alta baixa alta Perda Neutra
MS/MS Ion trap alta alta baixa média MSn
LC/TOF/MS alta média alta média Massa exta e sensibilidade
LC/QTOF/MS média alta alta média Massa exata dos fragmentos
Instrumento Características Vantagens Desvantagens
Q-TOF MS/MS Massa exata dos fragmentos de íons
Muito sensível em MS/MS para quase todos íons precursores selecionados
Full scan do íon precursor
Muito caroNão é possível perda neutra
Triplo Quadrupolo MS/MS
Perda Neutra Sensível para quantificação no modo MRM Não dá a massa exata
Quadrupolo Ion Trap MSn Sensível no scan de fragmentos facilmente deduziveis
Não é possível Perda Neutra e
MRM
68
Seletividade superior em LC-MS comparada com LC-UV
Será a técnica LC-TOF-MS a melhor solução?
69
Figures of Merit for Mass Analyzers
Type m/z range Resolving Power CostDouble Focusing 2-5000 20,000 $$$$Single Focusing 1-1400 1500 $$$
Quadrupole 1-4000 1000 $$Ion trap 10-4000 1000 $$Time of flight 1-100,000 30,000
$$$Fourier Transform 18-10,000 >100,000 $$$$
70
Development of Methods for the determination of cyanotoxins in surface and drinking water by using SPE and LC/MS-MS
LC/MS/MS SRM chromatograms of a standard solution of MC-RR, MC-YR, MC-LR, MC-LA and NOD (20 mg/L)T. Triantis et al., Toxicon, 2010, 55, 979-989
71
LC–MS/MS analyses of strigolactones in root exudates
Yoneyama K et al. Plant Cell Physiol 2010;51:1095-110372
73
Water soluble vitaminsLC-MS/MS
74